    ASR-2000 Series User Manual 
116 
T Ipeak, hold 
The T Ipeak, hold function sets the hold time for the peak current 
measurement. After the output is turned on, the ASR-2000 will 
delay starting the peak current measurement by this hold time. 
Concept in 
diagram 
Output on
Ipeak Hold time
I
Hold the peak current 
measurement value
T
 Note 
  The hold peak current will be updated when new 
measurement is greater than the previous value. 
By contrast, it will not update when new 
measurement is less than the hold peak current. 
  Ipeak hold time begins counting only when the 
output function is turned on. 
  Ipeak measurement is not executed during the 
set Ipeak hold time. 
Steps 
1.  Press the Menu key. The Menu 
setting will appear on the display. 
2.  Use the scroll wheel to go to item 2, MISC 
Configuration and press Enter. 
3.  Go to the T Ipeak, hold(msec) setting using the 
scroll wheel and press Enter. Set the time and 
press Enter again to confirm.  
T Ipeak 
1 ~ 60,000 ms 