User Manual
    ASR-2000 Series User Manual 
106 
Using External Control I/O 
Overview 
The External Control I/O is primarily used to 
control ASR-2000 externally by using the logic 
signal. More than that, it is able to monitor 
Sequence function status remotely with ease. 
Specification 
Control 
input 
  High level: +2.2 V or higher 
  Low level: +1.0 V or lower 
  Non-destructive maximum input: 
+7 V / -5 V 
  Input Impedance: Pulled up to 
+5 V with 47 kΩ 
Status 
output 
  Output level: 0 / +5 V 
  Output Impedance: 100 Ω 
Pin Assignment 
Check the table below for definition of each pin. 
Pin No. 
I/O 
Function 
Remark 
1 
Output 
 Power source on/off status 
 0: OFF, 1: On 
2 
Output 
 The output on/off status 
 0: OFF, 1: On 
3 
Output 
 0: OFF, 1: On 
4 
Output 
 Software busy status 
 0: Normal, 1: Busy 
5 
Output 
 Sequence sync output 0 
6 
Output 
 Sequence sync output 1 
7 
Output 
 Undefined output 0 
8 
Output 
 Undefined output 1 
9 
GND 
10 
Input 
 Undefined input 0 
11 
Input 
 Output off 
 Falling edge detection 
12 
Input 
 Output on 
 Falling edge detection 
13 
Input 
 Sequence start 
 Falling edge detection 
14 
Input 
 Sequence stop 
 Falling edge detection 
15 
Input 
 Sequence hold 
 Falling edge detection 
16 
Input 
 Sequence branch 1 
 Falling edge detection 
17 
Input 
 Sequence branch 2 
 Falling edge detection 