User Manual User guide

APS-1102A User Manual
APS-1102A
9-6
9.2 Current Limiter
a) Output peak current limiter When limiter is operating, output voltage is clipped.
Positive current setting range: 100 V range: +10.0 A to +42.0 A (Initial value: +42.0 A)
200 V range: +5.0 A to +21.0 A (Initial value: +21.0 A)
Negative current setting range: 100 V range: 42.0 A to 10.0 A (Initial value: 42.0 A)
200 V range: 21.0 A to 5.0 A (Initial value: 21.0 A)
Setting resolution: 0.1 A
b) Output RMS current limiter When limiter is operating, suppresses output voltage.
Current setting range: 100 V range: 1.0 A to 10.5 A (Initial value: 10.5 A)
200 V range: 1.0 A to 5.3 A (Initial value: 5.3 A)
Setting resolution: 0.1 A
9.3 Setting Range Limits
When signal source mode is INT or ADD only
a) Voltage setting limit
Positive voltage setting range: 100 V range: +0.1 V to +220.0 V (Initial value: +220.0 V)
200 V range: +0.1 V to +440.0 V (Initial value: +440.0 V)
Negative voltage setting range: 100 V range: 220.0 V to 0.1 V (Initial value: 220.0 V)
200 V range: 440.0 V to 0.1 V (Initial value: 440.0 V)
Setting resolution: 0.1 V
b) Frequency setting limit
(Lower limit upper limit)
Upper limit setting range: 1.0 Hz to 550.0 Hz (Initial value: 550.0 Hz)
Lower limit setting range: 1.0 Hz to 550.0 Hz (Initial value: 1.0 Hz)
Setting resolution: 0.1 Hz