APS-1102A User Manual
APS-1102A
6-46
IDN?
This returns the model name, etc. The response message includes the “manufacturer, model name,
serial number, and firmware version”.
Syntax: IDN?
Parameters: None
Query example: IDN?
Response format: <SRD>
Response example: GW Instek, APS-1102A, 000001, Ver1.00”
OPC
When all operations
have been completed, this sets the standard event register OPC bit (BIT0).
OPC? returns 1 when all operations have been completed. However, the standard event register
OPC bit is not cleared to zero when OPC? is executed.
Syntax: OPC
Parameters: None
Command example: OPC
Query example: OPC?
Response format: <NR1>
Response example: 1
RCL
This recalls status information stored to memory by SAV. The selection range for the recall
memory is 1 to 30. However, recall is disabled when output is on. ([1, Invalid with output on]
error occurs.)
Syntax: RCL<NR1>
Parameters: 1 to 30
Command example: RCL 1
Response format: None
Response example: None