User Manual User guide
APS-1102A User Manual
APS-1102A
6-38
STATus:OPERation:NTRansition
STATus:OPERation:PTRansition
This sets the operation transition filter (OPTF). The setting ranges from 0 to 20482.
When the NTR filter is set to 1, OPER is set to 1 when OPCR changes from 1 to 0.
When the PTR filter is set to 1, OPER is set to 1 when OPCR changes from 0 to 1.
When the NTR and PTR filters are both set to 1, OPER is set to 1 when OPCR changes.
When the NTR and PTR filters are both set to 0, changes in OPCR are not propagated to OPER.
Syntax: STATus:OPERation:NTRansition<NR1>
STATus:OPERation:PTRansition<NR1>
Parameters: 0 to 20482
Command example: STAT:OPER:NTR 16384
STAT:OPER:PTR 0
Query example: STAT:OPER:NTR?
STAT:OPER:PTR?
Response format: <NR1>
Response example: 16384 (when NTR is queried)
0 (when PTR is queried)
STATus:WARNing:CONDition?
The warning condition register (WRCR) value is returned. The response ranges from 0 to 29947.
Syntax: STATus:WARNing:CONDition?
Parameters: None
Query example: STAT:WARN:COND?
Response format: <NR1>
Response example: 16384