User Manual User guide

APS-1102A User Manual
APS-1102A
6-34
MEASure[:SCALar]:POWer:AC:REACtive?
This returns the reactive power in var units. The reactive power measurement full scale is 1400 var.
If the measured value exceeds the measurement range, 9999 var is returned as a measurement
error.
Syntax: MEASure[:SCALar]:POWer:AC:REACtive?
Parameters: None
Query example: MEAS:POW:AC:REAC?
Response format: <NR1>
Response example: 269
MEASure[:SCALar]:POWer:AC[:REAL]?
The effective power is returned in W units. The effective power measurement full scale is 1200 W.
If the measured value exceeds the measurement range, 9999 W is returned as a measurement error.
Syntax: MEASure[:SCALar]:POWer:AC[:REAL]?
Parameters: None
Query example: MEAS:POW:AC?
Response format: <NR1>
Response example: 249
MEASure[:SCALar]:VOLTage[:AC]?
This returns the output voltage RMS value (AC+DC RMS value) in Vrms units. For the 100 V
range, the output voltage RMS value measurement full scale is 250.0 Vrms, and for the 200 V
range, it is 500.0 Vrms. If the measured value exceeds the measurement range, 999.9 Vrms is
returned as a measurement error.
Syntax: MEASure[:SCALar]:VOLTage[:AC]?
Parameters: None
Query example: MEAS:VOLT?
Response format: <NR2>
Response example: 100.4