d) MEASure subsystem
MEASure[:SCALar]:CURRent[:AC]?
This returns the output current RMS value (AC+DC RMS value) in Arms units. The output current
RMS value measurement full scale is 15.00 Arms. If the measured value exceeds the measurement
range, 99.99 Arms is returned as a measurement error.
Syntax: MEASure[:SCALar]:CURRent[:AC]?
Parameters: None
Query example: MEAS:CURR?
Response format: <NR2>
Response example: 12.85
MEASure[:SCALar]:CURRent:HIGH?
This returns the highest current peak value (AC+DC peak value) in Apk units. The highest current
peak value measurement full scale is 45.0 Apk. If the measured value exceeds the measurement
range, 99.9 Apk is returned as a measurement error.
Syntax: MEASure[:SCALar]:CURRent:HIGH?
Parameters: None
Query example: MEAS:CURR:HIGH?
Response format: <NR2>
Response example: 25.9
MEASure[:SCALar]:CURRent:LOW?
This returns the lowest current peak value (AC+DC peak value) in Apk units. The lowest current
peak value measurement full scale is 45.0 Apk. If the measured value exceeds the measurement
range, 99.9 Apk is returned as a measurement error.
Syntax: MEASure[:SCALar]:CURRent:LOW?
Parameters: None
Query example: MEAS:CURR:LOW?
Response format: <NR2>
Response example: 35.7