User Manual User guide

APS-1102A User Manual
APS-1102A
6-26
[SOURce:]SEQuence:TPARameter
This sets the step transition parameters.
The target sequence step must be set in advance by the SEQ:STEP. ( See
[SOURce:]SEQuence:STEP on page 6-27.)
Use SYST:TUN ( See “ SYSTem:TUNit on page 6-43) in advance to set the unit for the
sequence step execution time or query the unit.
When in AC-EXT, AC-SYNC, AC+DC-EXT, or AC+DC-SYNC mode, the step transition
parameter cannot be set. ([3, Invalid in this mode] error occurs.)
Syntax: [SOURce:]SEQuence:TPARameter<NR2>, <NR1>......
First parameter (step time [s]/[ms])
Type <NR2>
Range 0.0001 to 999.9999 [s]
0.1 to 999999.9 [ms]
Second parameter (step end phase setting mode)
Type <NR1>
Range 0 (disable) or 1 (enable)
Third parameter (step end phase [])
Type <NR2>
Range 0.0 to 359.9
Fourth parameter (step termination)
Type <NR1>
Range 0 (continue), 1 (end), or 2 (hold)
Fifth parameter (jump step number)
Type <NR1>
Range 0 (next step), 1 to 255
Sixth parameter (jump times)
Type <NR1>
Range 0 (infinite times), 1 to 999
Seventh parameter (branch 0)
Type <NR1>
Range 0 (disabled), 1 to 999
Eighth parameter (branch 1)
Type <NR1>
Range 0 (disabled), 1 to 999
Command example: SEQ:TPAR 2, 1, 180, 0, 0, 1, 0, 0
Query example: SEQ:TPAR?
Response example: 2, 1, 180, 0, 0, 1, 0, 0