9.5 Measurement Functions
Output voltage measurement
RMS value: RMS value of AC+DC
Full scale: 225.0 Vrms: 100 V range
450.0 Vrms: 200 V range
Display resolution: 0.1 Vrms
Measurement accuracy: ±0.5% of the full scale (at 23 ±5°C)
Average value: Average value of AC+DC (measures DC component)
Full scale: ±225.0 V: 100 V range
±450.0 V: 200 V range
Display resolution: 0.1 V
Measurement accuracy: ±0.5% of the full scale (at 23 ±5°C)
Peak value: Positive peak voltage and negative peak voltage are displayed
separately
Full scale: +225 Vpk / 225 Vpk: 100 V range
+450 Vpk / 450 Vpk: 200 V range
Display resolution: 1 Vpk
Measurement accuracy: ±3% of the full scale (at 23 ±5°C, sine wave)
Output current measurement
RMS value: RMS value of AC+DC
Full scale: 15.00 Arms
Display resolution: 0.01 Arms
Measurement accuracy: ±1.0% of the full scale (at 23 ±5°C)
Average value: Average value of AC+DC (measures DC component)
Full scale: ±15.00 A
Display resolution: 0.01 A
Measurement accuracy: ±1.0% of the full scale (at 23 ±5°C)
Peak value: Positive peak current and negative peak current are displayed
separately
Full scale: +45.0 Apk / 45.0 Apk
Display resolution: 0.1 Apk
Measurement accuracy: ±3% of the full scale (at 23 ±5°C, sine wave)
Peak value hold: + peak current and peak current |maximum value
Full scale: 45.0 Apk
Display resolution: 0.1 Apk
Measurement accuracy: 3% of the full scale (at 23 5°C, sine wave)