6.3 Command Descriptions
APS-1102 Programmable AC/DC Power Source 6-33
g) OUTPut subsystem
OUTPut[:STATe]
This controls on/off settings of the output relay.
Syntax: OUTPut[:STATe] <bool>
Parameters: OFF|0 / ON|1
Command example: OUTP ON
Query example: OUTP?
Response format: <NR1>
Response example: 1
h) INPut subsystem
INPut:GAIN
This sets the external input gain used in the external mode (AC-EXT or AC+DC-EXT) and
internal + external mode (AC-ADD or AC+DC-ADD). For the 100 V range, the external input
gain setting ranges from 0.0 to 200.0 and for the 200 V range, it ranges from 0.0 to 400.0.
When in AC-EXT, AC-ADD, AC+DC-EXT, or AC+DC-ADD mode, the output waveform setting
cannot be made. ([3, “Invalid in this mode”] error occurs.)
Syntax: INPut:GAIN <NR2>
Parameters: 0.0 to 200.0 (0.0 to 400.0)
Command example: INP:GAIN 10.0
Query example: INP:GAIN?
Response format: <NR2>
Response example: 10.0