User Manual Manual

6.3 Command Descriptions
APS-1102 Programmable AC/DC Power Source 6-28
MEASure[:SCALar]:VOLTage[:AC]?
This returns the output voltage RMS value (AC+DC RMS value) in Vrms units. The output
voltage RMS value measurement full scale is 450.0 Vrms. If the measured value exceeds the full
scale, 999.9 Vrms is returned as a measurement error.
Syntax: MEASure[:SCALar]:VOLTage[:AC]?
Parameters: None
Query example: MEAS:VOLT?
Response format: <NR2>
Response example: 100.4
MEASure[:SCALar]:VOLTage:HIGH?
This returns the positive voltage peak value (AC+DC peak value) in Vpk units. The positive
voltage peak value measurement full scale is 450 Vpk. If the measured value exceeds the full
scale, 999 Vpk is returned as a measurement error.
Syntax: MEASure[:SCALar]:VOLTage:HIGH?
Parameters: None
Query example: MEAS:VOLT:HIGH?
Response format: <NR1>
Response example: 100
MEASure[:SCALar]:VOLTage:LOW?
This returns the negative voltage peak value (AC+DC peak value) in Vpk units. The negative
voltage peak value measurement full scale is 450 Vpk. If the measured value exceeds the full
scale, 999 Vpk is returned as a measurement error.
Syntax: MEASure[:SCALar]:VOLTage:LOW?
Parameters: None
Query example: MEAS:VOLT:LOW?
Response format: <NR1>
Response example: 99