6.3 Command Descriptions
APS-1102 Programmable AC/DC Power Source 6-27
MEASure[:SCALar]:POWer:AC:APParent?
This returns the apparent power in VA units. The apparent power measurement full scale is 1400
VA. If the measured value exceeds the full scale, 9999 VA is returned as a measurement error.
Syntax: MEASure[:SCALar]:POWer:AC:APParent?
Parameters: None
Query example: MEAS:POW:AC:APP?
Response format: <NR1>
Response example: 367
MEASure[:SCALar]:POWer:AC:PFACtor?
This returns the output power factor. The power factor measurement range is 0.00 to 1.00. When
the measured value exceeds the measurement range, 9.99 is returned as a measurement error.
The power factor is calculated as “effective power/apparent power”.
Syntax: MEASure[:SCALar]:POWer:AC:PFACtor?
Parameters: None
Query example: MEAS:POW:AC:PFAC?
Response format: <NR2>
Response example: 0.68
MEASure[:SCALar]:POWer:AC:REACtive?
This returns the reactive power in Var units. The reactive power measurement full scale is 1400
Var. If the measured value exceeds the full scale, 9999 Var is returned as a measurement error.
Syntax: MEASure[:SCALar]:POWer:AC:REACtive?
Parameters: None
Query example: MEAS:POW:AC:REAC?
Response format: <NR1>
Response example: 269
MEASure[:SCALar]:POWer:AC[:REAL]?
The effective power is returned in W units. The effective power measurement full scale is 1200 W.
If the measured value exceeds the full scale, 9999 W is returned as a measurement error.
Syntax: MEASure[:SCALar]:POWer:AC[:REAL]?
Parameters: None
Query example: MEAS:POW:AC?
Response format: <NR1>
Response example: 249