User Manual Manual
6.3 Command Descriptions
APS-1102 Programmable AC/DC Power Source 6-24
d) MEASure subsystem
MEASure[:SCALar]:CURRent[:AC]?
This returns the output current RMS value (AC+DC RMS value) in Arms units. The output
current RMS value measurement full scale is ±15.00 Arms. If the measured value exceeds the full
scale, 99.99 Arms is returned as a measurement error.
Syntax: MEASure[:SCALar]:CURRent[:AC]?
Parameters: None
Query example: MEAS:CURR?
Response format: <NR2>
Response example: 12.85
MEASure[:SCALar]:CURRent:HIGH?
This returns the positive current peak value (AC+DC peak value) in Apk units. The positive
current peak value measurement full scale is 45.0 Apk. If the measured value exceeds the full
scale, 99.9 Apk is returned as a measurement error.
Syntax: MEASure[:SCALar]:CURRent:HIGH?
Parameters: None
Query example: MEAS:CURR:HIGH?
Response format: <NR2>
Response example: 25.9
MEASure[:SCALar]:CURRent:LOW?
This returns the negative current peak value (AC+DC peak value) in Apk units. The negative
current peak value measurement full scale is 45.0 Apk. If the measured value exceeds the full
scale, 99.9 Arms is returned as a measurement error.
Syntax: MEASure[:SCALar]:CURRent:LOW?
Parameters: None
Query example: MEAS:CURR:HIGH?
Response format: <NR2>
Response example: 25.9