Programmable AC/DC Power Source APS-1102 USER MANUAL GW INSTEK PART NO.
This manual contains proprietary information, which is protected by copyrights. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company. The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice. Good Will Instrument Co., Ltd. No.
Preface Thank you for purchasing the APS-1102 programmable AC/DC power source. To ensure safe and proper use of this electric equipment, please read first Safety Precautions on the following pages. Caution Symbols Used in This Manual The following caution symbols are used in this manual. Be sure to observe these caution symbols and their contents to ensure the safety of the user and avoid damage to the equipment.
Safety Precautions To ensure safe use, be sure to observe the following warnings and cautions. Good Will shall not be held liable for damages that arise from a failure to observe these warnings and cautions. This product is a Class 1 product (with protective conductor terminal) that conforms to the JIS and IEC insulation standards. Be sure to observe the contents of this user manual. This user manual contains information for the safe operation and use of this product.
Prevention of electrical shock due to output voltage The maximum output of the APS-1102 is 400 V. Carefully operate the APS-1102 to avoid electrical shock. Directly touching the output or changing cable connections while output is on may cause electrical shock. Safety-related symbols The general definitions of the safety-related symbols used on this product and in the user manual are provided below.
Contents Page Preface ........................................................................................................... i Safety Precautions ....................................................................................... ii 1. OVERVIEW ............................................................................................... 1-1 1.1 1.2 1.3 1.4 1.5 2. PREPARATIONS BEFORE USE ............................................................... 2-1 2.1 2.2 2.3 2.4 2.5 3. General ..
Contents 4.1.2 Selecting an output mode (AC+DC-INT mode) .................................................... 4-3 4.1.3 Setting the output voltage range .......................................................................... 4-5 4.1.4 Setting the output voltage .................................................................................... 4-7 4.1.5 Setting output frequency of superimposed AC ................................................... 4-9 4.1.6 Setting superimposed AC waveform ....
Contents 5.3.5 Output initial phase settings .............................................................................. 5-19 5.3.6 Waveform settings .............................................................................................. 5-19 5.3.7 Output peak current limiter settings .................................................................. 5-20 5.3.8 Output average current limiter settings ............................................................. 5-21 5.4 MEASURE Screen ....
Contents 7. TROUBLESHOOTING .............................................................................. 7-1 7.1 Protection Function.................................................................................................. 7-2 7.2 Error Messages and Responses.............................................................................. 7-4 7.2.1 Error at power-on .................................................................................................. 7-5 7.2.
Figures Figure 1-1. Page Block Diagram............................................................................................................ 1-6 Figure 2-1. Ambient Temperature and Humidity Ranges ............................................................ 2-4 Figure 2-2. Power Source Inlet .................................................................................................... 2-7 Figure 2-3. Operation Check .........................................................................
Figures Figure 5-1. LCD Screen (When Normal) ....................................................................................... 5-3 Figure 5-2. LCD Screen (When Warning Is Displayed) ............................................................... 5-3 Figure 5-3. Menu Tree ................................................................................................................... 5-7 Figure 5-4. Modification Box ..............................................................................
Tables Page Table 1-1. List of Functions (1/2) ............................................................................................... 1-4 Table 1-2. List of Functions (2/2) ............................................................................................... 1-5 Table 2-1. Panel Settings for Operation Check ......................................................................... 2-9 Table 3-1. Settings in Memory (1/2) ...............................................................
Tables Table 5-11. SET Menu Items and Output Modes ........................................................................ 5-15 Table 5-12. Output Modes........................................................................................................... 5-16 Table 5-13. Output Voltage Settings .......................................................................................... 5-18 Table 5-14. Output Frequency Setting ......................................................................
Tables Table 7-14. When Fault Is Suspected (Problem Related to Measurement Functions) ............. 7-17 Table 7-15. When Fault Is Suspected (Problem Related to Sequence Function) ..................... 7-17 Table 7-16. When Fault Is Suspected (Problem Related to Memory Function) ........................ 7-17 Table 7-17. When Fault Is Suspected (Problem Related to Limiter Setting Range Limit) ....... 7-18 Table 7-18. When Fault Is Suspected (Other Problems) ..........................................
1. O VERVIEW 1.1 1.2 1.3 1.4 1.5 General ......................................................................... 1-2 Features ........................................................................ 1-2 Applications ................................................................. 1-3 List of Functions .......................................................... 1-4 Operation Principles ....................................................
1.1 General 1.1 General The APS-1102 programmable AC/DC power source is a power source that can output AC and DC power, features a compact design for convenient desktop use, and provides a wealth of measurement functions. Rated output voltage is 100 Vrms (100 V range) or 200 Vrms (200 V range), with maximum output capacity of 1 kVA (during AC 200 V input). It also features eight output modes, with DC output, external input amplification, and line-synchronized output.
1.3 Applications Enables AC-superimposed DC output When in the AC+DC mode, an AC wave (sine wave, square wave, or arbitrary waveform) can be superimposed on the DC output. Also supports capacitor input load (up to crest factor 4 is supported) Enables peak current output of up to four times as large as the rated current (RMS value). Can be used as an external input amplifier When “internal + external” is selected, an internal signal source can be added to the external signal input.
1.4 List of Functions 1.4 List of Functions The main functions of the APS-1102 are listed below. Table 1-1.
1.4 List of Functions Table 1-2. List of Functions (2/2) Function External External control I/O control USB interface Description External control operation modes: Enable, disable Control input: Input level: High level: +4.0 V or higher Low level: +1.0 V or lower Non-destructive max.
1.5 Operation Principles 1.5 Operation Principles Figure 1-1 shows a block diagram of the APS-1102. PANEL <2> CONTROLLER USB SYSTEM CONTROL CONTROL I/O DISPLAY OSC AMP CONTROL EXT SIG IN 2.2V MAX Zin=10k EXT SYNC IN TTL <3> ISOLATION INVERTER <1> PFC INVERTER <4> AMP INVERTER SUB-DCPS CIRCUIT ~LINE 100V-230V 50Hz/60Hz 1.4kVA MAX POWER ISO CIRCUIT (PRI) PFC CIRCUIT LINE FILTER ISO CIRCUIT (SEC) OUTPUT AC250V MAX 1kVA MAX POWER PROTECT AMP OUTPUT 400Vpk MAX 1kVA MAX Figure 1-1.
2. P REPARATIONS BEFORE USE 2.1 2.2 2.3 2.4 2.
2.1 Checking Before Use 2.1 Checking Before Use Before installing and using the APS-1102, make sure that it has not been damaged during shipment, and check that all the accessories and all parts of the main unit are included. If anything is missing, contact the Good Will distributor from which the product was purchased.
2.2 Installation Environment 2.2 Installation Environment Note the following precautions to ensure the safe use and reliability of this product. Installation sites Do not set the product on its top or side when installing it on a floor or desktop. Make sure that the four rubber feet on the bottom of the product are set evenly on a flat surface when placed on a floor or desktop.
2.2 Installation Environment Conditions of installation site The APS-1102 should be used indoors, and at an elevation of 2,000 meters. The APS-1102 uses a fan for forced-air cooling. To allow for ample air flow, be sure to maintain a gap of at least 50 cm between the air inlets and outlets along the sides and rear of this product and walls or other obstructions. Install this product in a location that meets the following conditions for temperature and humidity ranges.
2.3 Grounding and Power Supply Connection Do not install the APS-1102 in the following locations: Location with flammable gas An explosion may occur. Never install and use this product in such a location. Outdoors, or location exposed to direct sunlight or near a fire or heat source The full performance of this product may not be obtained, or failure may occur. Location with corrosive gas, moisture, or high humidity This product may become corroded or fail.
2.3 Grounding and Power Supply Connection Frequency range: 50 Hz/60 Hz ±2 Hz (single phase) Power consumption: 1.
2.3 Grounding and Power Supply Connection Connect the power in this order. 1. Make sure that the power source voltage to be used is within the specified range. 2. Set the power switch to off. 3. Insert the power cord into the inlet on the rear panel. Inlet Figure 2-2. Power Source Inlet 4. Insert the power cord plug into a 3-prong power source outlet (use the supplied power source cord set 1).
2.4 Simple Operation Checks 2.4 Simple Operation Checks The following describes simple methods for checking newly purchased products or products that have been in long-term storage. ! WARNING This product contains high-voltage parts. Never remove the cover. All internal inspections of this product are to be performed only by service technicians qualified by Good Will.
2.4 Simple Operation Checks 2. Panel settings See “3.4 Main Operation Example”, for descriptors Set the APS-1102 as shown below. of the settings methods. Table 2-1. Panel Settings for Operation Check Item Setting Output mode AC-INT mode Output voltage range 100 V AC output voltage 100 Vrms Selected measurement display RMS 3. Press the OUTPUT key to turn on output. Output on/off LED is on. OUTPUT Output is off. OUTPUT Output is on.
2.5 Calibration 2.5 Calibration To calibrate the APS-1102, contact Good Will or an authorized distributor.
3. P ANEL AND BASIC OPERATIONS 3.1 Panel Components and Operations ····································· 3-3 3.1.1 Operation panel ···························································· 3-3 3.1.2 Front panel ···································································· 3-4 3.1.3 Rear panel ····································································· 3-5 3.2 Display and Initial Settings at Power-on ······························ 3-6 3.
3.1 Panel Components and Operations 3.1 Panel Components and Operations 3.1.1 Operation panel Numbers shown after names in this figure correspond to item numbers in the detailed descriptions below. LCD 5.1 CANCEL key Output ON/OFF LED 3.4.8 ENTER key 5.2 MODIFY dial 5.2 DIGIT key OUTPUT key 3.4.8 UP/DOWN keys Shortcut keys 5.2 5.2 5.2 5.2.6 Figure 3-1.
3.1 Panel Components and Operations 3.1.2 Front panel Numbers shown after names in the following two figures correspond to item numbers in the detailed descriptions below. Power switch 3.4.1 Exhaust vent 8.2 Output outlets (universal type) 3.3.1 Figure 3-2.
3.1 Panel Components and Operations 3.1.3 Rear panel External signal input/external sync signal input connector 3.3.4 Exhaust vent Output terminals LINE INPUT USB connector 3.3.1 2.3 3.3.2 External control I/O connector 3.3.3 Figure 3-3.
3.2 Display and Initial Settings at Power-on 3.2 Display and Initial Settings at Power-on Display at power-on When the power switch is set to on, a fault check is performed automatically, and when that ends normally operation mode is set. The settings shown on the panel are those that were shown the last time the power source was turned off. When a newly purchased unit is turned on for the first time, the initial settings are displayed. See “4.
3.2 Display and Initial Settings at Power-on Table 3-2.
3.3 I/O Terminals 3.3 I/O Terminals 3.3.1 Output terminals (front and rear) Connection to output outlet (front) The front panel is equipped with universal type output outlets that accommodate power plug types from various countries. These outputs are isolated from the power inputs and case (ground connection). These outlets are used only for AC power. Use the rear output terminal for DC or AC+DC. Polarity follows the Japanese standard.
3.3 I/O Terminals Connection to output terminals (rear) Outputs are isolated from the input power lines and the case (ground connection). When connecting to an output terminal, attach a round crimp contact with a sleeve that has sufficient capacity for the output current. DC output is based on “Lo”, so that the + polarity setting sets “Hi = positive”, and the - polarity setting sets “Hi = negative”. Output terminal Figure 3-5.
3.3 I/O Terminals Connect cables to the output terminal block as described below. Outputs are isolated from the input power lines and case. 1. Remove a screw attached to the output terminal block. 2. Insert the screw through the round crimp contact at the end of the cable. 3. Reattach the screw into the OUTPUT terminal while holding the crimp contact in place. 4. Turn the screw until it is securely fastened. The recommended fastening torque is 1.2 (Nm). 5.
3.3 I/O Terminals 3.3.2 USB connector Use a commercial (off-the-shelf) USB cable to connect the rear USB connector to a PC. Attach the accompanying ferrite core to the USB cable to reduce the leakage electric wave. See “6. USB INTERFACE”, for description of using a USB connection for remote control from a PC. USB connector Figure 3-7. USB Connector ! CAUTION The USB cable should be a commercial Type A (male) to Type B (male) cable.
3.3 I/O Terminals 3.3.3 External control I/O The APS-1102 is equipped with an (8-bit) external control I/O connector to enable external control using digital signal input and status output using digital signals. The connector is a D-sub 25-pin type. External control I/O connector Figure 3-8. External Control I/O Connector The signals corresponding to the pins in this connector are listed below in “Table 3-3 External Control I/O Connectors”.
3.3 I/O Terminals Table 3-3. External Control I/O Connectors Pin No.
3.3 I/O Terminals 3.3.4 External signal input/external sync signal input terminal When using an externally input signal as the signal source with external synchronization, connect to the external signal input/external sync signal input terminals (EXT SIG IN/EXT SYNC IN) on the rear panel. For this connection, use a coaxial cable with a BNC connector. The specifications for these external signal input/external sync signal input terminals are listed below.
3.4 Main Operation Example 3.4 Main Operation Example Main operations are described below in the following order. Power-on 3.4.1 Power on/off 3.4.2 Setting output mode Set output mode 3.4.3 Setting output voltage range Set output voltage range 3.4.4 Setting waveform Set waveform 3.4.5 Setting output voltage Set output voltage 3.4.6 Setting output frequency Set output frequency Use current limiter and setting range limiter 3.4.7 Using limiter functions 3.4.
3.4 Main Operation Example 3.4.1 Power on/off When the POWER switch is set to “|”, the power goes on, and when it is set to “”, the power goes off. POWER switch Figure 3-10. POWER Switch a) Power-on Set the POWER switch to “|”. Startup screen is displayed.
3.4 Main Operation Example Main operations screen is displayed. Output mode 3.4.2 Output voltage range Output voltage 3.4.3 3.4.5 Output frequency Waveform Limiter function 3.4.6 3.4.4 3.4.7 Select measurement display 3.4.9 Figure 3-11. Example of Main Operations Screen (in AC-INT Mode) The main operations screen that appears immediately after the power-on shows the same settings that were displayed the last time the power was turned off.
3.4 Main Operation Example 3.4.2 Setting output mode The APS-1102 provides the eight output modes listed in Table 3-4. The output mode cannot be changed while output is on. Before changing the output mode, set output to off. Table 3-4. Output Mode List Operation Signal Source Mode Mode AC Reference INT (internal) AC-INT “3.4.2 Setting output mode” EXT (external) AC-EXT “4.5 Control Using External Control I/O Connector” “4.
3.4 Main Operation Example Operation steps Select output mode by selecting 1. Press the or in the SET menu. key to move the cursor to the icon. Cursor CANCEL 2. Press the ENTER ENTER key to display the selection box. CANCEL 3. Press the or CANCEL 4. Press the ENTER CANCEL Selection box ENTER key to move the cursor to the mode to be set. ENTER key to set the selected mode.
3.4 Main Operation Example 3.4.3 Setting output voltage range Select 100 V or 200 V. The output voltage range cannot be changed while output is on. The setting range varies according to the output voltage range. Table 3-5.
3.4 Main Operation Example 2. Press the ENTER key to display the selection box. Selection box CANCEL 3. Press the ENTER or CANCEL 4. Press the ENTER key to move the cursor to the output voltage range to be selected. ENTER key. Then the icon appears in the status window. Status window CANCEL ⇒ The ENTER icon goes out in a few seconds, then the display will change to the selected output voltage range. Changing the output voltage range takes about eight seconds.
3.4 Main Operation Example 3.4.4 Setting waveform The APS-1102 allows to select a sine wave, square wave, or arbitrary waveform (16 types in all) as the waveform to be output, by using the control panel. Arbitrary waveforms are generated using data transfers via the USB interface. See “4.6 Output of Arbitrary Waveforms”, for details of arbitrary waveforms. Table 3-6 lists the waveforms that can be selected. Table 3-6.
3.4 Main Operation Example 3. Press the or CANCEL 4. Press the ENTER key to move the cursor to the waveform to be output. ENTER key to set the selected waveform. CANCEL ENTER 3.4.5 Setting output voltage For AC-INT mode, the output voltage settings listed below can be selected. Table 3-7. Output Voltage Settings Output Mode Setting AC-INT AC (AC mode) voltage 100 V range 200 V range Setting Range Resolution SIN/SQU 0.0 to 140.0 Vrms 0.1 Vrms ARB1 to ARB 16 0.0 to 400.0 Vp-p 0.
3.4 Main Operation Example Operation steps To set the output voltage, set 1. Press the or in the SET menu. key to move the cursor to the icon. Cursor CANCEL 2. Press the ENTER ENTER key to display the modification box. Modification box CANCEL 3. Press the ENTER and keys to move the cursor to the digit to be changed, turn the MODIFY dial and set a numerical value. The set value will be reflected in output immediately. Decrease 4.
3.4 Main Operation Example 3.4.6 Setting output frequency The AC output frequency can be set in the following range. Table 3-8. Output Frequency Setting Setting Range Resolution Setting Accuracy 1.0 Hz to 550.0 Hz 0.1 Hz ±100 ppm (1.0 Hz to 550.0 Hz, 23 ±5°C) See “3.4.7 Using limiter functions”, when using an output frequency setting range limit. Operation steps To set the output frequency, set a numerical value labeled in the SET menu. 1. Press the icon.
3.4 Main Operation Example 3.4.7 Using limiter functions a) Using current limiter The APS-1102 includes an output peak current limiter function and an output average current limiter function. Limiter settings are retained separately for 100 V and 200 V ranges. Setting range limits can be set for output voltage and output frequency. See “5.3.7 Set Output peak current limiter settings”. See “5.3.8 Set Output average current limiter settings”. Table 3-9.
3.4 Main Operation Example Operation steps Select the icon in the SET menu corresponding to the desired setting, then set a numerical value. Output peak current limiter (positive): Output peak current limiter (negative): Output average current limiter: 1. Press the or CANCEL 2. Press the ENTER ENTER or or icon. Cursor key to display the modification box. CANCEL 3.
3.4 Main Operation Example b) Using setting range limiter The APS-1102 internal mode (AC-INT, AC+DC-INT) and internal + external mode (AC-ADD, AC+DC-ADD) include setting range limiter functions for output voltage and output frequency. See “5.5.5 Setting range limiter (LIMIT)”, for details. Table 3-10. Voltage and Frequency Setting Range Limiter Setting Setting Range Resolution Initial Value 100 V range +5.0 V to +200.0 V 0.1 V +200.0 V 200 V range +10.0 V to +400.0 V 0.1 V +400.
3.4 Main Operation Example 3. Press the ENTER key to display the LIMIT screen . CANCEL ENTER 4. Press the or key to move the cursor to the item to be set. Cursor CANCEL 5. Press the ENTER ENTER key to display the modification box. Modification box CANCEL 6. Press the ENTER and keys to move the cursor to the digit to be changed, turn the MODIFY dial and set a numerical value. The set value will be reflected in output immediately. Decrease 7.
3.4 Main Operation Example 3.4.8 Output on/off OUTPUT To turn the APS-1102 output on or off, simply press the key. When output is on, the output on/off LED goes on, and it goes off when the output is off. OUTPUT OUTPUT Output off ! Output on/off LED on Output on WARNING To ensure safety, make sure the power is off before connecting outputs. ! CAUTION Before turning off the power, make sure that the output on/off LED is off.
3.4 Main Operation Example 3.4.9 Using measurement functions The APS-1102 is equipped with the following measurement functions.
3.4 Main Operation Example Operation steps To select a measurement display, select “MEASURE” in the MISC menu 1. Press the or CANCEL key to move the cursor to the . icon. ENTER Cursor 2. Press the ENTER Press the or CANCEL 3. Press the key to display the MISC menu. ENTER CANCEL key to move the cursor to “MEASURE”. ENTER key to display the selection box.
3.4 Main Operation Example 4. Press the or CANCEL key to move the cursor to the desired measurement display. ENTER HC1 to HC4 are for measuring harmonic current. See “4.3 Measurement of Harmonic Current”. 5. Press the ENTER key to set the selected measurement display. Selected measurement display is shown. CANCEL ENTER 6. Press the TOP MENU key or CANCEL key to return to the SET menu screen.
3.5 Output Characteristics 3.5 Output Characteristics The following charts show output voltage vs. output current of the APS-1102 (resistance load, typical character). Figures 3-13 and 3-14 show charts when AC is input, and Figures 3-15 and 3-16 show when DC is Output current (Arms)/Maximum output current (Arms) 100 (%) input. 100.0 75.0 57.
Output current (A)/Maximum output current (A) 100 (%) 3.5 Output Characteristics 100.0 75.0 41.0 Maximum output power = 750 W Output range: Rated voltage/Maximum voltage --------------------------------------------------------------------- 100 V range: 100 V/10 A 200 V range: 200 V/5 A 75.0 100.0 183 200 Output voltage (V)/Rated output voltage (V) × 100 (%) Figure 3-15. Output Voltage vs.
4. A DVANCED OPERATION EXAMPLE 4.1 Use as a DC Power Source ··················································· 4-2 4.1.1 Connecting output terminal to load during DC output ··········································································· 4-2 4.1.2 Selecting an output mode (AC+DC-INT mode) ··········· 4-3 4.1.3 Setting the output voltage range ·································· 4-5 4.1.4 Setting the output voltage ············································ 4-7 4.1.
4.1 Use as a DC Power Source 4.1 Use as a DC Power Source When using the APS-1102 as a DC voltage source, select DC - Internal signal mode (AC+DC-INT), then enter the settings listed in Table 4-1. AC components (sine wave, square wave, and arbitrary waveform) can be superimposed on the DC power source. A protection diode may be required to be connected, depending on the load, such as a capacitor or inductor. Table 4-1.
4.1 Use as a DC Power Source Once a diode has been connected, a load that does not consume power keeps its voltage charged. The voltage of the load may be slightly higher due to the overshoot voltage and should be carefully monitored. APS-1102 Block diode Hi Capacitor Lo Figure 4-1. Connection of Block Diode b) Inductive load such as inductor An inductive load generates a back EMF when current suddenly stops flowing, such as when the output is turned off.
4.1 Use as a DC Power Source 2. Press the ENTER CANCEL 3. Press the key to display the selection box. Selection box ENTER or key to move the cursor to “AC+DC-INT”. Cursor CANCEL 4. Press the ENTER CANCEL ENTER key to set the AC+DC-INT mode.
4.1 Use as a DC Power Source 4.1.3 Setting the output voltage range Select 100 V or 200 V as the output voltage range. As is shown in Table 4-2, the range that can be set for the output voltage differs according to the selected output voltage range. Table 4-2. Setting Range Options for Various Output Voltage Ranges Setting range Setting 100 V range 200 V range DC voltage 200.0 to +200.0 400.0 to +400.0 SIN/SQU 0.0 to 140.0 0.0 to 280.0 AC voltage ARB1 to ARB16 0.0 to 400.0 0.0 to 800.
4.1 Use as a DC Power Source 3. Press the or CANCEL 4. Press the ENTER CANCEL key to move the cursor to 100 V or 200 V. ENTER key. Then the icon appears in the status window . ENTER The icon goes out in a few seconds, then the output voltage range changes to the selected range. Changing the output voltage range takes about eight seconds.
4.1 Use as a DC Power Source 4.1.4 Setting the output voltage The ranges that can be set are listed below. Table 4-3. Output Voltage Setting Ranges When AC+DC Mode Is Selected Setting resolution Setting range 0.1 V 100 V range 200.0 to +200.0 V DC voltage 0.1 V 200 V range 400.0 to +400.0 V 0.1 Vrms SIN/SQU 0.0 to 140.0 Vrms 100 V range 0.1 Vp-p ARB1 to ARB16 0.0 to 400.0 Vp-p AC voltage 0.1 Vrms SIN/SQU 0.0 to 280.0 Vrms 200 V range 0.1 Vp-p ARB1 to ARB16 0.0 to 800.
4.1 Use as a DC Power Source 3. Press the and keys to move the cursor to the digit to be changed, turn the MODIFY dial and set a numerical value. The set value will be reflected in output immediately. Decrease 4. Press the CANCEL or ENTER Increase key to close the modification box.
4.1 Use as a DC Power Source 4.1.5 Setting output frequency of superimposed AC To set the output frequency, set a numerical value under in the SET menu. Operation steps 1. Press the or key to move the cursor to the icon. Cursor CANCEL 2. Press the ENTER ENTER key to display the modification box. Modification box CANCEL 3. Press the ENTER and keys to move the cursor to the digit to be changed, turn the MODIFY dial and set a numerical value.
4.1 Use as a DC Power Source 4.1.6 Setting superimposed AC waveform Selects among 16 types of sine waves (SIN), square waves (SQU), and arbitrary waveforms (ARB). Arbitrary waveforms are generated using data transfers via the USB interface. See “4.6 Output of Arbitrary Waveforms”, for details of arbitrary waveforms. To select a waveform, select in the SET menu. Operation steps 1. Press the or CANCEL 2. Press the ENTER CANCEL 3. Press the CANCEL key to move the cursor to the icon.
4.1 Use as a DC Power Source 4. Press the ENTER CANCEL key to set the selected waveform. ENTER 4.1.7 Use measurement function Measured values are displayed in the measured value window MEASURE screen, which enables the output status to be monitored. a) Measured value window Figure 4-3. Example of Measured Value Window (During AC+DC Mode) When a measurement is out of range, “-----” is displayed in the numerical value field.
4.1 Use as a DC Power Source b) Select measurement display Select one form among RMS, AVG, or PEAK as the measured current and voltage values to be displayed. To select a measurement display, select “MEASURE” in the MISC menu . Operation steps 1. Press the CANCEL or key to move the cursor to the icon. ENTER Cursor 2. Press the ENTER Press the CANCEL 3. Press the CANCEL key to display the MISC menu. or key to move the cursor to “MEASURE”. ENTER ENTER key to display the selection box.
4.1 Use as a DC Power Source 4. Press the or key to move the cursor to the desired measurement display. Select “AVG” to measure only the DC power source. CANCEL ENTER 5. Press the CANCEL ENTER key to set the selected measurement display. ENTER Although harmonic current measurement (HC1 to HC4) is also displayed in the selection box, this cannot be selected when in AC+DC mode. When in AC internal mode (AC-INT), harmonic current measurement can be selected only when the frequency setting is 50.
4.2 Measurement of Inrush Current 4.2 Measurement of Inrush Current 4.2.1 Inrush current In such an electrical product as contains a capacitor input type rectifier circuit for the input power line, a current that is larger than the rated current may flow for a short time immediately after the power source switch is turned on. This current is called an “inrush current.
4.2 Measurement of Inrush Current 4.2.2 Set phase at output on To set the phase at output on, select “Phase at output on” in the SET menu and set a numerical value. Before changing the phase at output on, set output off. Operation steps 1. Press the or CANCEL 2. Press the ENTER icon. Cursor ENTER key to display the modification box. CANCEL 3.
4.2 Measurement of Inrush Current 4.2.3 Set measurement display to peak value Set the displayed current measurement value to the peak value (PEAK). To change the displayed current measurement value, select “MEASURE” in the MISC menu . Operation steps 1. Press the CANCEL or key to move the cursor to the icon, then press the ENTER key. ENTER Cursor The MISC menu is displayed. 2. Press the or key to move the cursor to “MEASURE”, then press the ENTER key. The selection box is displayed.
4.2 Measurement of Inrush Current 4. Press the TOP MENU key or the CANCEL key to return to the SET menu screen. The measured value window shows the peak value set in the measurement display. Peak value Peak hold value TOP MENU START SEQUENCE STOP HOLD LOCAL 4.2.4 Reset the peak current hold value The APS-1102 includes a function that resets the peak current hold value.
4.2 Measurement of Inrush Current 2. Press the or key to move the cursor to “RESET Ipkh…”, then press the ENTER key. The EXEC box is displayed. CANCEL EXEC box ENTER 3. When “EXEC” is displayed, press the ENTER key. The peak current hold value is reset. CANCEL ENTER 4. Press the TOP MENU key or CANCEL key to return to the SET menu screen. The peak current hold value shown in measured value window becomes 0 Apk.
4.3 Measurement of Harmonic Current 4.3 Measurement of Harmonic Current Switching power sources are widely used in both consumer and industrial electrical products. The capacitor input type rectifier circuits that are used with such input power lines have disadvantages such as significant distortion of input current and a large amount of harmonic current.
4.3 Measurement of Harmonic Current 3. Press the the ENTER or key to move the cursor to the harmonic order to be displayed, then press key. CANCEL ENTER When HC1 to HC4 have been selected as the displayed measurements, the following harmonic current measured values are displayed. The displayed harmonic order (XXth) changes according to these ranges, and 10 lines of measurement data are displayed in each range.
4.4 Using the Sequence Function 4.4 Using the Sequence Function 4.4.1 Sequence operations When the sequence function of the APS-1102 is used, the pattern to be output can be freely programmed. For example, if a series of voltage change patterns that will be used in voltage variation tests for car electronics are programmed in advance, the output will change according to the sequence, so the test can be completed in one series of operations.
4.4 Using the Sequence Function a) Sequence A sequence is comprised of a combination of at least two steps (the minimum unit is one step). b) Step Step execution parameters and step transition parameters are specified for each step. When sequence execution is not performed (i.e., during normal output mode), these parameters are specified in the same way. Thereafter, the normal output mode is called normal output step. In the descriptions below, the step parameters are enclosed in brackets ([]).
4.4 Using the Sequence Function The step transition parameters specify the next step and its timing. Table 4-5. Step Transition Parameters [Step time] [Step end phase] [Step end] [Jump step] [Jump times] [Branch step] Specified in time Sets step end phase as enabled or disabled. The following parameters are provided to specify the action at the end of the step.
4.4 Using the Sequence Function An example of basic step transitions is shown below. Output parameters Steps 1, 2 Step 3 Normal output step Step 1 Step 2 Time Step 3 Figure 4-5. Example of Basic Step Transitions c) Sequence mode Figure 4-6 shows a mode transition diagram for sequence operations. Output on Idle mode (normal output) Run mode (sequence execution) Hold mode (pause) Figure 4-6.
4.4 Using the Sequence Function d) Sequence control The following describes how sequence operations are controlled. Sequence control operations are indicated in triangular brackets (< >). Table 4-7. Sequence Control Sets Run mode. When changing from Idle mode, the operation goes to step 1 when is set, and the sequence is started. When changing from Hold mode, the sequence is resumed from where it was stopped when was set.
4.4 Using the Sequence Function e) Sequence loop Either an entire sequence or a block of any number of steps within a sequence can be used as a sequence loop. The loop is implemented based on the [jump step] and [jump times] parameters specified for each step. The jump counter, which counts the number of jumps, functions within each step to retain a count, so multiple loops can be implemented. Figure 4-9 below illustrates the loop specification method and the step transitions.
4.4 Using the Sequence Function f) Step synchronized output The code that is specified in the step being executed is output to the external control I/O connector. However, “0” is output during a normal output step. Step synchronized output Bit 0 1 0 [Step synchronized output] Step 1 Step 2 Step 3 Step 4 Step 5 0 1 0 0 1 Time Figure 4-10. Example of Step Synchronized Output g) Step end phase setting When the step end phase is set as enabled, the step execution time is as described below.
4.4 Using the Sequence Function 4.4.2 Sequence settings Sequence settings are made as described below. Input per step via control panel Programmed from PC, with USB transfer to APS-1102 Sequence data is edited and transferred using accompanying control software See “4.4.3 Programming sequences” below, for further description of the program messages for sequence operations. See the control software user’s manual, for description of the control software.
4.4 Using the Sequence Function 3. Press the ENTER CANCEL 4. Press the key to display the SEQUENCE screen. ENTER or key to select the operation to be performed. To control sequence operations: select “CONTROL…”. ( See “4.4.4 Control of sequence operations”.) Note This cannot be selected when output is off. To program a sequence: Select “PROGRAM”. ( See “4.4.3 Programming sequences”.) To control sequence memory: Select “MEMORY…”. ( See “4.4.5 Clear sequence memory”.) CANCEL 5.
4.4 Using the Sequence Function 4.4.3 Programming sequences When “PROGRAM” is selected in the SEQUENCE screen, the PROGRAM screen is displayed. The sequence is programmed according so the various settings in the PROGRAM screen. To change to the SEQUENCE screen, the output mode must first be switched to AC-INT or AC+DC-INT. Enter step settings per step. Table 4-8 to Table 4-10 list the setting items in the PROGRAM screen.
4.4 Using the Sequence Function Table 4-8. Setting Items in PROGRAM Screen (1/3) Setting Name Item STEP Step number TIME Step time DC voltage value DC VOLT (AC+DC-INT only) DC operation type AC voltage value AC VOLT AC operation type Operation Specifies the step number to be set. Setting range: 1 to 255 Specifies the continuation time for the specified step. Setting range: 0.0001 s to 999.9999 s Setting resolution: 0.0001 s (= 0.
4.4 Using the Sequence Function Table 4-9. Setting Items in PROGRAM Screen (2/3) Setting Name Item AC frequency FREQ AC frequency operation type Waveform WAVE Waveform operation type Phase PHS Phase operation type Step synchronized output SYNC CODE Step synchronized output operation type Step end phase enable/disable STOP PHS Step end phase Operation Sets the AC frequency. Setting range: 1.0 Hz to 550.0 Hz Setting resolution: 0.1 Hz Sets the method used to change to the specified frequency.
4.4 Using the Sequence Function Table 4-10. Setting Items in PROGRAM Screen (3/3) Setting Name Item STEP TERM Step end JUMP STEP Jump step JUMP NUM Jump times BRANCH0 Branch step 0 BRANCH1 Branch step 1 Operation Sets operation after end of step. Select one of the following three options. CONT: Sequence continuation STOP: Change to Idle mode HOLD: Change to Hold mode Sets the step number to jump to after the end of the previous step.
4.4 Using the Sequence Function 1. Open the SEQUENCE screen and display the PROGRAM screen. See “4.4.2 Sequence settings”, for steps for displaying the PROGRAM screen. 2. Press the or key to move the cursor to “TIME”. Cursor CANCEL 3. Press the ENTER ENTER key to display the modification box. Modification box CANCEL 4. Press the ENTER and keys to move the cursor to the digit to be changed, turn the MODIFY dial and set a numerical value. The set value will be reflected in output immediately.
4.4 Using the Sequence Function Operation step <2> The settings for DC voltage operation type (AC+DC mode only), AC voltage operation type, frequency operation type, waveform (WAVE), waveform operation type, phase operation type, step synchronized output (SYNC CODE), step synchronized output operation type, step end phase enable/disable (STOP PHS), and step end (STEP TERM) are made by selecting settings for the various items in the order described below.
4.4 Using the Sequence Function 4.4.4 Control of sequence operations The CONTROL item in the SEQUENCE screen is used to control sequence operations. A sequence can be started only when output is on. Consequently, the sequence control selection box can be selected only when output is on. Table 4-11 lists items in the CONTROL selection box. Table 4-11. Items in CONTROL Selection Box Selection Box Items START STOP HOLD BRAN0 BRAN1 Operation Starts sequence operation. Switches to Run mode.
4.4 Using the Sequence Function b) When operating with the shortcut keys Operation step The control of sequence operations, can also be performed with the shortcut keys at the bottom of the control panel. The START, STOP, and HOLD keys under “SEQUENCE” correspond to items in the CONTROL selection box. Shortcut keys cannot be used to control branching. Sequence shortcut keys 5.2.6 Figure 4-12.
4.4 Using the Sequence Function 4.4.5 Clear sequence memory Data is saved to sequence memory automatically. Separate sequence data can be saved for 100 V range and 200 V range in AC-INT mode and for 100 V range and 200 V range in AC+DC-INT mode. When clearing the sequence memory, use “MEMORY…” in the SEQUENCE screen. The operation steps are shown below. Operation steps 1. Open the SEQUENCE screen and display the MEMORY EXEC box. See “4.4.
4.4 Using the Sequence Function 4.4.6 Example of sequence operation settings The following describes the steps for setting a new sequence operation, based on the sequence transition example described in Table 4-12. The time unit is seconds (s). ( See “5.6.5 Time unit setting” to change the time unit.) Output voltage 100.0 V Jump to step 2 “KEEP” setting is valid. 75.0 V 50.0 V Setp 5 setting value Jump to step 6 upon external trigger input. <3> <2> <1> 25.0 V 0.0 V Step 1 (0.1 s) Step 2 (0.
4.4 Using the Sequence Function The operation steps are described below. Operation steps using control panel 1. Open the PROGRAM screen. The steps for opening the PROGRAM screen are described in “4.4.2 Sequence settings”. Setting step 1 2. Press the key to move the cursor to “DC VOLT”. CANCEL 3. Press the ENTER ENTER key to display the modification box. CANCEL 4. Press the ENTER and keys to move the cursor to the 10’s place, then turn the MODIFY dial to set “2”.
4.4 Using the Sequence Function 6. Press the key to move the cursor to the line displaying “CONST” (fourth line from the top). CANCEL 7. Press the ENTER ENTER key to display the selection box. Selection box CANCEL 8. Press the or CANCEL 9. Press the ENTER ENTER key to move the cursor to “SWEEP”. ENTER key to confirm the setting. The settings for all other items are the same as the default settings, so they are not changed.
4.4 Using the Sequence Function Setting step 2 10. Press the 11. Press the 12. Press the or ENTER key to move the cursor to “STEP”. key to display the modification box. and keys to move the cursor to 1’s place, then turn the MODIFY dial to set “2”. The set value will be reflected in output immediately. 13. Press the CANCEL 14. Press the 15. Press the 16. Press the or ENTER key to close the modification box. key to move the cursor to “TIME”. ENTER key to display the modification box.
4.4 Using the Sequence Function 4.4.7 Execution of sequence operations Once sequence programming has been completed, start the sequence. The execution steps for the sequence operations are described below. a) When using the MISC menu Operation steps 1. Press the TOP MENU key (shortcut key) to return to the top menu. In the top menu, “SET” is shown in the upper left corner of the menu window. The sequence is executed only when output is on. Make sure output is on beforehand. TOP MENU START 2.
4.4 Using the Sequence Function 4. Press the ENTER key to display the selection box. CANCEL ENTER 5. Press the key to move the cursor to “ON”. CANCEL 6. Press the ENTER ENTER key to set output on, at which time the output on/off LED goes on. OUTPUT OUTPUT Output off Output on 7. Select “SEQUENCE”, then press the CANCEL Output on/off LED is on ENTER key to display the SEQUENCE screen .
4.4 Using the Sequence Function 8. Press the key to select “CONTROL”, then press the ENTER key to display the selection box. Selection box CANCEL ENTER 9. Select “START”, then press the ENTER key to start the sequence. b) When using keys on the control panel Operation steps 1. Press the TOP MENU key to return to the top menu, check the settings in the menu, then press the OUTPUT key on the control panel. This set output on, at which time the output on/off LED goes on.
4.4 Using the Sequence Function c) Operations after starting sequence When a sequence is started, the status icon ( ) is displayed in the status window to indicate that the sequence is being executed. A message box is also displayed to indicate the step number being executed. After returning to the SET screen, if the sequence is executed, the current step settings will not be shown in the SET screen. When the sequence is completed, the settings from the final step will be shown.
4.5 Control Using External Control I/O Connector 4.5 Control Using External Control I/O Connector Digital signals that are input to the external control I/O connector can be used to control sequences and the output on/off setting. The sequence step synchronized output signals are output from the external control I/O connector. The status of the APS-1102 can be monitored by reading the output signals from the external control input/output connector. See “3.3.
4.6 Output of Arbitrary Waveforms 4.6 Output of Arbitrary Waveforms Arbitrary waveforms are generated by transferring data from an external PC to arbitrary waveform memory via a USB interface. Arbitrary waveforms can be edited and transferred using the accompanying control software. See the control software user’s manual, for description of this control software. When output is on, arbitrary waveforms cannot be changed. Make sure that output is off before changing arbitrary waveforms.
4.6 Output of Arbitrary Waveforms 2. Press the key to move the cursor to the arbitrary waveform to be selected (“ARB1” or to “ARB16”), then press the ENTER key. This sets the selected arbitrary waveform.
4.7 Synchronization of Output to External Signal 4.7 Synchronization of Output to External Signal The APS-1102 is equipped with an externally synchronized oscillation function. The software PLL function synchronizes the output frequency to the frequency of the external synchronization TTL signal. The output frequency can be synchronized to any frequency from 40 to 500 Hz. The synchronization phase difference cannot be set.
4.7 Synchronization of Output to External Signal 4.7.1 External signal synchronization This function synchronizes the output frequency to a signal that is input to the rear panel external sync signal input (BNC connector). The output frequency can be synchronized to any frequency from 40 to 500 Hz. Operation steps Select in the SET menu to select external synchronization mode, then set the sync signal source to external sync signal “EXT”. 1. Make sure that output is off.
4.7 Synchronization of Output to External Signal 4. Press the ENTER CANCEL 5. Press the key to select the mode to be set. ENTER or key to move the cursor to the icon, then press the ENTER key to display the selection box. CANCEL 6. Press the Selection box ENTER or key to move the cursor to “EXT”, then press the ENTER key. This sets the external sync signal source to external sync signal “EXT”.
4.8 Using Memory Functions 4.7.2 Line synchronization This function synchronizes the output frequency to the frequency of the AC power source line. The output frequency can be synchronized to power source line frequency (50 Hz/60 Hz). Operation steps Select in the SET menu to select external synchronization mode, then set the sync signal source to line synchronization (“LINE”). 1. Perform the same operations as were described above for steps 2 to 5 in “4.7.
4.8 Using Memory Functions Memory control is performed via the “MEMORY” screen. After selecting an operation listed in Table 4-14, use the selection box to specify the target memory number. When a memory number is specified, the main settings stored in that memory are displayed in the left side of the menu window. Table 4-14. Menu Items in MEMORY Screen Indication on Control Panel STORE... RECALL... CLEAR... Type Stores settings to specified memory number. Recalls settings from specified memory number.
4.8 Using Memory Functions Operation steps Select and then “MEMORY” in the MISC menu to store, recall, or clear memory or to set the target memory number. 1. Press the or CANCEL key to move the cursor to the icon. ENTER Cursor 2. Press the ENTER Press the or CANCEL 3. Press the key to display the MISC menu. ENTER CANCEL key to move the cursor to “MEMORY”. ENTER key to display the MEMORY screen.
4.8 Using Memory Functions 4. Press the or CANCEL 5. Press the ENTER CANCEL 6. Press the CANCEL key to select the operation to be performed. ENTER key to display the selection box. ENTER or key to move the cursor to the target memory number. ENTER 7. Check the settings shown in the left side of the menu window, and if they are OK press the ENTER key. This performs the operation specified in step 4 for the specified memory number.
4.9 Amplification of External Signal 4.9 Amplification of External Signal The APS-1102 can be used as an amplifier to which a signal is input from the external signal input terminals on the rear panel, by selecting the external signal source mode (AC-EXT or AC+DC-EXT). Connect the external signal source to the external signal input/external sync signal input terminal (EXT SIG IN/EXT SYNC IN) on the rear panel. For this connection, use a coaxial cable with a BNC connector.
4.9 Amplification of External Signal a) Possible voltage settings for output Operation steps Select the icon in the SET menu corresponding to the desired setting, then set a numerical value. “Select output mode”: Select external mode (AC-EXT or AC+DC-EXT). “Output voltage range”: Select 100 V or 200 V. “Set external input gain”: Set a numerical value. 1. Press the or key to move the cursor to he icon, then press the ENTER key. The selection box is displayed.
4.9 Amplification of External Signal the ENTER key. This sets the selected output voltage range. CANCEL 5. Press the ENTER or key to move the cursor to the icon, then press the ENTER key. The modification box is displayed. CANCEL 6. Press the ENTER and keys to move the cursor to the column to be changed, turn the MODIFY dial and set a numerical value. The set value is reflected immediately in the output. Decrease 7.
4.9 Amplification of External Signal b) Signal input The setup for external input is completed once the above operations and the settings are done. Now connect the signal source and input the signal. If the input signal exceeds the maximum voltage that can be output, the output overvoltage protection function will operate, in which case output may be turned off and the waveform may become distorted. Caution is therefore required concerning the input voltage level and the external input gain.
4.10 Adding External Signals and Internal Signals 4.10 Adding External Signals and Internal Signals When in the APS-1102 internal + external mode (AC-ADD, AC+DC-ADD), the internal signal source signal can be added to the external signal input prior to output. When using a signal input from an external source as a signal source, connect to the rear panel external signal input/external sync signal input terminal (EXT SIG IN/EXT SYNC IN). For this connection, use a coaxial cable with a BNC connector.
5. M ENUS 5.1 Screen Configuration ···························································· 5-3 5.1.1 Status icons ·································································· 5-5 5.1.2 Warnings and error messages ······································ 5-6 5.1.3 Sequence display ·························································· 5-6 5.2 Basic Operations ·································································· 5-7 5.2.
5.6.1 5.6.2 5.6.3 5.6.4 5.6.5 5.6.6 5.6.7 5.6.8 5.6.
5.1 Screen Configuration 5.1 Screen Configuration The LCD on the control panel splits in multiple windows, as is shown in Figure 5-1 and Figure 5-2. A B D C Figure 5-1. LCD Screen (When Normal) E Figure 5-2.
5.1 Screen Configuration The following information is displayed in various parts of the screen. A: Status window: This window shows the status of various devices. B: Message box: When an error has occurred, a message box appears near the bottom of the status window, and the cause is displayed. The message box is closed automatically after 1.5 seconds. It is also used to indicate an overload, as well as the progress of a sequence. C: Measured value window: This window displays measured values.
5.1 Screen Configuration 5.1.1 Status icons The status window displays status icons that indicate the status of the APS-1102. These status icons are listed below in Table 5-1. Table 5-1. Status Icons List Icon Item Description Battery almost depleted When a backup memory checksum error occurs, this icon flashes to indicate that the battery is almost depleted. Unlock This icon goes on when external line synchronization is disabled.
5.1 Screen Configuration 5.1.2 Warnings and error messages When a warning or error occurs, an error message is displayed in a message box or message display window. See “7 TROUBLESHOOTING”, for further description of error messages. 5.1.3 Sequence display The APS-1102 is equipped with a function that indicates the progress of the sequence being executed. The information listed in Table 5-2 is displayed in the status window. Table 5-2.
5.2 Basic Operations 5.2 Basic Operations The APS-1102 basic operations are described below. 5.2.1 Menus The APS-1102 uses a hierarchical menu system for setting various parameters. Menu operations are performed using the menu window in the LCD screen. Figure 5-3 illustrates the APS-1102 menu tree. (1) SET (2) MISC (3) SEQUENCE (4) PROGRAM (5) MEMORY (6) REMOTE (7) SYSTEM (8) INFORMATION (9) LIMIT Figure 5-3. Menu Tree The items in these menus are listed in Table 5-3 to Table 5-9.
5.2 Basic Operations Table 5-3. Items in SET Menu Icon Item Reference Select output mode See “5.3.1 Output mode setting”. Select output voltage range See “5.3.2 Output voltage range setting”. Set AC voltage to be output See “5.3.3 Setting output voltage ”. Set DC voltage to be output See “5.3.3 Setting output voltage ”. Set frequency to be output See “5.3.4 Setting output frequency”. Set phase when output is on See “5.3.5 Output initial phase settings”. Set waveform to be output See “5.3.
5.2 Basic Operations Table 5-4. Items in MISC Menu Setting Name Setting Reference SEQUENCE Open sequence menu See “5.5.1 Sequence (SEQUENCE)”. MEMORY Open memory menu See “5.5.2 Memory (MEMORY)”. REMOTE Open remote menu SYSTEM Open system menu LIMIT Open setting range limit menu MEASURE Select measurement display RESET Ipkh Reset peak current hold value OUTPUT Select output on/off See “5.5.3 Remote (REMOTE)”. See “5.5.4 System (SYSTEM)” and “5.6 System Menu”. See “5.5.
5.2 Basic Operations Table 5-5. Items in SEQUENCE Menu Setting Name CONTROL... PROGRAM STEP TIME DC VOLT AC VOLT FREQ WAVE PHS SYNC CODE STOP PHS STEP TERM JUMP STEP JUMP NUM BRANCH0 BRANCH1 MEMORY...
5.2 Basic Operations Table 5-7. Items in REMOTE Menu Setting Name LOCAL... Setting Switch to local mode Reference See “5.5.3 Remote (REMOTE)”. Table 5-8. Items in SYSTEM Menu Setting Name KEYLOCK BEEP Keylock Beep CONTRAST LCD contrast COLOR LCD display color TIME UNIT Time unit INIT OUT Output on/off at power-on EXT CONTROL RESET... INFORMATION Setting External control input enable/disable Reset Display system information Reference See “5.6.1 Keylock”. See “5.6.2 Beep sound”. See “5.6.
5.2 Basic Operations 5.2.2 Navigating the menu tree Use the UP and DOWN keys to move among items displayed in the menu windows. Press the ENTER key to go to a lower-level menu in the menu tree and press the CANCEL key to go to the higher-level menu. Operation steps The specific operation steps are as follows. 1. Use the UP and DOWN keys to move the menu cursor displayed in the menu window, then select a menu item that is marked along the right with a ▶ symbol. 2.
5.2 Basic Operations 5.2.3 Numerical value input operations When input of numerical values is required, the modification box is displayed. When the modification box is opened, the digit cursor is displayed in one column. Use the DIGIT key to move the digit cursor and use the MODIFY key to increase or decrease values. Each time the MODIFY key is used to change a value in the modification box, the actual setting is updated immediately.
5.2 Basic Operations 5.2.5 EXEC input The EXEC box is displayed for items to be actually executed by pressing the ENTER key. Only one target item is displayed in the EXEC box, and the cursor is shown on that item. When the ENTER key is pressed, the item is actually executed, and the EXEC box is closed. If the EXEC box is closed by pressing the CANCEL key, nothing is executed. Figure 5-6. EXEC Box 5.2.6 Shortcut keys The APS-1102 shortcut keys can be used to perform the following operations. Table 5-10.
5.3 SET Menu 5.3 SET Menu The SET menu in the menu window is used to set basic functions. The available settings differ according to the output mode. Table 5-11. SET Menu Items and Output Modes Item INT Output Mode AC AC+DC EXT ADD SYNC INT EXT ADD SYNC Selection of output mode See “5.3.1 Output mode setting”. Selection of output voltage range See “5.3.2 Output voltage range setting”. Setting of AC voltage to be output See “5.3.3 Setting output voltage”.
5.3 SET Menu 5.3.1 Output mode setting Table 5-12 lists the eight output modes that can be selected (based on combinations of two operation modes and four signal source modes). The mode cannot be changed when output is on. Turn output off before entering settings. Table 5-12.
5.3 SET Menu Figure 5-7 shows a block diagram of the APS-1102 signal sources. SYSTEM CONTROL EXT INPUT GAIN EXT SIG IN EXT SYNC EXT ADD EXT To AMP CONTROL PLL LINE LINE INT ADD SYNC SYNC INT OSC Figure 5-7. Block Diagram of Signal Sources Operation steps Select in the SET menu. See “3.4.2 Setting output mode”, for how to select the output mode. 5.3.2 Output voltage range setting Select 100 V or 200 V as the output voltage range. The output voltage range cannot be changed when output is on.
5.3 SET Menu 5.3.3 Setting output voltage The output voltage settings listed below can be made for the two operation modes. Table 5-13. Output Voltage Settings Operation Mode Setting AC mode (AC-INT, AC-ADD, AC-SYNC) AC voltage DC voltage AC+DC mode (AC+DC-INT, AC+DC-ADD, AC+DC-SYNC) AC voltage 100 V range 200 V range 100 V range 200 V range 100 V range 200 V range Setting Range 0.0 to 140.0 Vrms 0.0 to 400.0 Vp-p 0.0 to 280.0 Vrms 0.0 to 800.0 Vp-p 200.0 to +200.0 V 400.0 to +400.0 V 0.0 to 140.
5.3 SET Menu 5.3.4 Setting output frequency This sets the AC voltage frequency. When either AC-SYNC or AC+DC-SYNC mode is selected, the AC voltage frequency cannot be set. Table 5-14. Output Frequency Setting Setting Range Resolution Setting Accuracy 1.0 Hz to 550.0 Hz 0.1 Hz ±100 ppm(1.0 Hz to 550.0 Hz,23 ±5C) See “3.4.7 Using limiter functions”, when using a setting range limit for output frequency. Operation steps in the SET menu.
5.3 SET Menu 5.3.7 Output peak current limiter settings The positive value and the negative value of the peak current limiter can be set independently. If the output peak current exceeds the set current, the output voltage is clipped. Setting range limits can be set for output voltage and output frequency. See “5.5.5 Setting range limiter (LIMIT)”. Table 5-16.
5.3 SET Menu 5.3.8 Output average current limiter settings This limits the output average current. When the output average current exceeds this setting, the output voltage is suppressed. Table 5-17. Output Average Current Limiter Setting Range Setting Output average current limiter Setting Range Resolution Initial Value 100 V range 1.0 A to 10.5 A 0.1 A 10.5 A 200 V range 1.0 A to 5.3 A 0.1 A 5.3 A Operation steps Select in the SET menu and set numerical values. See “3.4.
5.4 MEASURE Screen 5.4 MEASURE Screen The MEASURE screen in the measured value window displays measured values for items selected as measurement display items (some of which depend on the output mode). The measured values for output voltage and output current are displayed as values set via “MEASURE” in the MISC menu. Item Table 5-18. Display Items in MEASURE Screen Select Unit Measurement Reference Display Voltage RMS value DC voltage value Vrms Vp-p RMS V AVG “5.4.
5.4 MEASURE Screen 5.4.1 Output voltage measurement When “MEASURE” has been selected in the MISC menu, RMS value (RMS), average DC (AVG), or peak value (PEAK) is displayed. The initial setting (factory setting) is to display the RMS value (RMS). See “3.4.9 Using measurement functions”. a) RMS value Full scale: RMS value of AC+DC ±225.0 Vrms: 100 V range ±450.0 Vrms: 200 V range Display resolution: 0.1 Vrms Measurement accuracy: ±0.
5.4 MEASURE Screen 5.4.2 Output current measurement When “MEASURE” has been selected in the MISC menu, RMS value (RMS), average DC (AVG), or peak value (PEAK) is displayed. The initial setting (factory setting) is to display the RMS value (RMS). See “3.4.9 Using measurement functions”. a) RMS value RMS value of AC+DC Full scale: 15.00 Arms Display resolution: 0.01 Arms Measurement accuracy: 1.
5.4 MEASURE Screen 5.4.3 Output power measurement The output power values described below are always displayed.
5.4 MEASURE Screen 5.4.6 Output harmonic current measurement Output harmonic current measurement is enabled only in the AC-INT mode and when the frequency setting is 50 or 60 Hz. When HC1, HC2, HC3, or HC4 is selected under “MEASURE” in the MISC menu, the harmonic measurement window is displayed. At that time, measurements other than output harmonic measurement are not performed. This measurement does not comply with the IEC standards. See “4.3 Measurement of Harmonic Current”.
5.5 MISC Menu 5.5 MISC Menu This menu is used to enter detailed settings for various functions. These settings can be entered regardless of whether output is on or off. 5.5.1 Sequence (SEQUENCE) A sequence menu is displayed. This menu is used to control sequences, set up programming, and clear sequence memory. a) CONTROL…(sequence control) Select among the sequence operation control items listed below. See “4.4.4 Control of sequence operations”. Table 5-19.
5.5 MISC Menu b) PROGRAM (sequence programming) This “PROGRAM” screen is used to program sequences. The items listed below in Table 5-20 to Table 5-22 are set for individual steps. Sequence programming is stored automatically, and the data for AC mode with a 100 V range or a 200 V range is stored separately from the data for AC+DC mode with a 100 V range or a 200 V range. See “4.4.3 Programming sequence”, for description of sequence programming operation steps. Table 5-20.
5.5 MISC Menu Table 5-21. PROGRAM Screen Items (2/3) Setting Name Item AC frequency FREQ AC frequency operation type Waveform WAVE Waveform operation type Phase PHS Phase operation type Step synchronized output SYNC CODE Step synchronized output operation type Step end phase STOP PHS enable/disable Step end phase Operation Sets the AC frequency. Setting range: 1.0 Hz to 550.0 Hz Setting resolution: 0.1 Hz Sets the method used to change to the specified frequency.
5.5 MISC Menu Table 5-22. PROGRAM Screen Items (3/3) Setting Name Item STEP TERM Step end JUMP STEP Jump step JUMP NUM Jump times BRANCH0 Branch step 0 BRANCH1 Branch step 1 Operation Sets operation after end of step. Select one of the following three options. CONT: Sequence continuation STOP: Change to Idle mode HOLD: Change to Hold mode Sets the step number to jump to after the end of the previous step.
5.5 MISC Menu 5.5.2 Memory (MEMORY) The “MEMORY” screen is used to control memory. Settings saved (stored) to store/recall memory No. 1 to No. 30 can be recalled and used. Recall of store/recall memory contents can be executed only when output is off. See “4.8 Using Memory Functions”, for description of store/recall memory. After selecting one of the following operations, use the selection box to specify the target memory number. STORE...: Stores settings to specified memory number. RECALL...
5.5 MISC Menu 5.5.3 Remote (REMOTE) The remote control connected via a USB interface can be used to switch to local mode for operation via the panel screen. When in remote control mode, the “RMT” icon is displayed. See “6 USB INTERFACE”. Operation steps 1. Select “LOCAL…”, then press the ENTER key. The EXEC box is displayed. Displayed during remote mode EXEC box CANCEL ENTER 2. Make sure “EXEC” has been selected, then press the ENTER key. This sets local mode.
5.5 MISC Menu 5.5.5 Setting range limiter (LIMIT) The setting range limiter indicates the LIMIT screen, enabling the setting range of output voltage and output frequency to be limited when the signal source mode is the internal signal source mode (AC-INT, AC+DC-INT) or internal + external signal source mode (AC-ADD, AC+DC-ADD), and the setting range of the output voltage to be limited when in the external sync mode (AC-SYNC, AC+DC+SYNC). The numeric values shown in Table 5-23 are set on this screen.
5.5 MISC Menu 5.5.6 Selection of measurement display (MEASURE) Select the measured values to be displayed, from the RMS value (RMS), average DC (AVG), peak value (PEAK), and harmonic current values (HC1 to HC4). When harmonic current value is selected, the RMS value, average DC, and peak value are not measured. When measurement of the RMS value, average DC, or peak value has been set, harmonic current is not measured. See “3.4.
5.5 MISC Menu 5.5.8 Output on/off OUTPUT The APS-1102 output on/off setting can be entered using the key on the front panel, but it can also be controlled via the MISC menu. See “3.4.8 Output on/off ”. Operation steps 1. Press the or CANCEL 2. Press the ENTER ENTER key to display the selection box. CANCEL 3. Press the ENTER key to move the cursor to “ON”. CANCEL 4. Press the key to move the cursor to “OUTPUT”.
5.6 System Menu 5.6 System Menu This menu is used to enter the following system settings for the APS-1102. KEYLOCK: Keylock BEEP: Beep CONTRAST: LCD contrast COLOR: LCD display color TIME UNIT: Time unit INIT OUT: Output on/off at power-on EXT CONTROL: External control input enabled/disabled RESET...: Reset INFORMATION: System information display “SYSTEM” screen display steps Select and then “SYSTEM” in the MISC menu. 1.
5.6 System Menu 3. Press the ENTER CANCEL 4. Press the key to display the SYSTEM screen . ENTER or key to select the operation to be performed.
5.6 System Menu 5.6.1 Keylock The APS-1102 is equipped with a function that prevents key input errors during operation. When “keylock” is on, all keys become inactive except for the keylock off key. Although it is possible to switch among menus when keylock is on, basically no changes can be made except for turning keylock off. However, output can still be turned off via the OUTPUT key as an emergency response measure. Operation steps 1. Press the or key to move the cursor to “KEYLOCK”.
5.6 System Menu 5.6.2 Beep sound When the beep sound is on, a beep sounds when an incorrect operation is made, a key is pressed, or a modify operation is made. When a protection function-related error occurs, an alarm sound will be output regardless of the beep sound setting. Operation steps 1. Press the or key to move the cursor to “BEEP”. Cursor CANCEL 2. Press the ENTER ENTER key to display the selection box. Selection box CANCEL 3.
5.6 System Menu 5.6.3 LCD contrast adjustment The LCD screen contrast level can be set to any of 100 levels to suit the operation environment. To open the level adjustment text box, press the ENTER key and CANCEL key at the same time. This enables the contrast to be adjusted even when the current contrast setting does not allow easy recognition of the LCD display contents. This also moves the digit cursor to the 10’s place. Operation steps 1. Press the or key to move the cursor to “CONTRAST”.
5.6 System Menu 5.6.4 LCD display color setting The display color of the LCD screen on the control panel can be set as either blue based (white text on blue background) or white based (blue text on white background). Operation steps 1. Press the or key to move the cursor to “COLOR”. Cursor CANCEL 2. Press the ENTER ENTER key to display the selection box. Selection box CANCEL 3. Press the ENTER or key to move the cursor to “BLUE” for blue based or “WHITE” for white based. CANCEL 4.
5.6 System Menu 5.6.5 Time unit setting The unit used to indicate the time parameter can be specified. Select either “s” (second) or “ms” (millisecond) for this time unit. Operation steps 1. Press the or key to move the cursor to “TIME UNIT”. Cursor CANCEL 2. Press the ENTER ENTER key to display the selection box. Selection box CANCEL 3. Press the or CANCEL 4. Press the ENTER ENTER key to move the cursor to the displayed unit. ENTER key to complete the setting.
5.6 System Menu 5.6.6 Output on/off setting at power-on The APS-1102 can be set to automatically turn output on at power-on. However, when restarting after a system lock, the APS-1102 will start with output off regardless of this setting. Operation steps 1. Press the or key to move the cursor to “INIT OUT”. Cursor CANCEL 2. Press the ENTER CANCEL 3. Press the key to display the selection box. ENTER Selection box ENTER or CANCEL 4.
5.6 System Menu 5.6.7 External control input enable/disable setting External control input can be set as enabled or disabled. However, the APS-1102 status will be output even when external control input is set as disabled. See “4.5 Control Using External Control I/O Connector”, for details of external control input. Operation steps 1. Press the or CANCEL key to move the cursor to “EXT CONTROL”. ENTER Cursor 2. Press the ENTER CANCEL 3. Press the ENTER or CANCEL 4.
5.6 System Menu 5.6.8 Reset function This function resets all basic settings to their initial (factory) settings. The APS-1102 cannot be reset when output is on. Always turn output off before resetting the APS-1102. See “3.2 Display and Initial Settings at Power-on”, for description of the settings after a reset. Operation steps 1. Press the or CANCEL key to move the cursor to “RESET…”. ENTER Cursor 2. Press the ENTER CANCEL key to display the EXEC box . ENTER EXEC box 3.
6. U SB INTERFACE 6.1 6.2 6.3 Setup for Use ········································································ 6-2 Command List ······································································ 6-3 Command Descriptions ························································ 6-9 6.3.1 Overview of Programming Language ··························· 6-9 6.3.2 Detailed command descriptions ································· 6-14 6.
6.1 Setup for Use 6.1 Setup for Use The APS-1102 can be controlled under the USB Test and Measurement Class (USBTMC) standard. Almost all control functions can be operated from the control panel, and internal status information such as settings and error information can be read. The USB connector is located on the rear panel. Attach the accompanying ferrite core to the USB cable to reduce the leakage electric wave. The USB interface is designed for use in a relatively favorable environment.
6.2 Command List 6.2 Command List The APS-1102 command list is shown in Table 6-1 to Table 6-9 and IEEE488.2 (SCPI) common commands supported by the APS-1102 are listed in Table 6-10. The symbols used in Table 6-1 to Table 6-10 are explained below. The lowercase letters in each keyword can be omitted. Square brackets indicate keywords that can be omitted. (Implicit keywords) A vertical bar ( | ) indicates that one of several keywords is selected. Table 6-1.
6.2 Command List Table 6-2.
6.2 Command List Table 6-4.
6.2 Command List Table 6-8. Command List (SYSTem Subsystem) Function Command Beep sound on/off SYSTem:BEEPer:STATe External control input enable/disable SYSTem:CONFigure:EXTIO Error query SYSTem:ERRor? Output at power-on SYSTem:PON[:OUTPut] Time unit for sequence step execution SYSTem:TUNit Clear warning SYSTem:WRELease Table 6-9.
6.2 Command List Table 6-10. Common Command List (Common Commands and Queries) Common Commands and Queries Name Function *CLS Clear command Clears standard event register, etc. *ESE Standard event enable command Sets standard event enable register. *ESR? Standard event register query Queries standard event register. *IDN? Identification query Queries device ID information.
6.2 Command List a) Input buffer The number of commands that can be sent at one time is limited by the input buffer capacity (10,000 characters). The transferred commands are first stored in the input buffer, then successively interpreted and executed. When commands are being interpreted and executed, an error occurs if an illegal command is found. That command and those following will not be executed.
6.3 Command Descriptions 6.3 Command Descriptions 6.3.1 Overview of Programming Language This products conforms to IEEE488.2 and SCPI (VERSION 1999.0). a) SCPI SCPI defines methods used for communications between computers and measuring instruments. Refer to the following for general information on SCPI. http://www.scpiconsortium.org Standard Commands for Programmable Instruments (SCPI) VERSION 1999.
6.3 Command Descriptions b) SCPI data format All command messages and acknowledge messages sent to the APS-1102 are in “ASCII” format. Table 6-11. Numerical Value Data Format Symbol Data Format Numerical values without a decimal point (decimal position is to the right of the smallest digit in value) Example: 273 Numerical value with explicit decimal point Example: .0273 Numerical value with explicit decimal point and exponent Example: 2.
6.3 Command Descriptions d) Subsystem commands Subsystem commands are commands used to execute the particular functions of devices. Subsystem commands correspond to a measuring instrument internal function groups, with several sets divided into groups. Each subsystem command has a tree structure, and is defined using colons (:) as path separators. 1) SCPI command tree Figure 6-2 shows a partial command tree of subsystem commands. See Figure 6-3 for an overview of the command tree. Figure 6-2.
6.3 Command Descriptions 2) Path separator The path separator (:) separates the current keyword from the keyword at the next lower level. Each time a colon (:) is detected in a command string, the current path is moved one level downward. When a colon appears at the start of a command string, it indicates that the current path is set to the root. This starting colon also can be freely omitted.
6.3 Command Descriptions 4) Implicit keywords Any keyword that is enclosed by square brackets ([ ]) is an implicit keyword that can be omitted. Abbreviated command string examples <1> to <5> below all send the same setting to the target device.
6.3 Command Descriptions 6.3.2 Detailed command descriptions The following describes various command functions and syntax. a) Meanings of symbols Angle brackets (< >): Parameter symbols are enclosed in angle brackets. For example, indicates the digital data format (NR1 type). A vertical bar ( | ) A parameter is used to alternative parameters. For example, “FIX|STEP” indicates that either “FIX” or “STEP” can be used as a parameter.
6.3 Command Descriptions [SOURce:]CURRent:LIMit:PEAK:HIGH This sets the upper limit of the output peak current (positive polarity). The output peak current limiter (positive polarity) setting ranges from 10.0 to 42.0 A (for 100 V range) or from 5.0 to 21.0 A (for 200 V range). Syntax: [SOURce:]CURRent:LIMit:PEAK:HIGH Parameters: 10.0 to 42.0 (5.0 to 21.0) Command example: CURR:LIM:PEAK:HIGH 10 Query example: CURR:LIM:PEAK:HIGH? Response format: Response example: 10.
6.3 Command Descriptions [SOURce:]FREQuency:LIMit:HIGH This sets the upper limit of the frequency that can be set. However, a value that is at or below the currently set frequency cannot be set. ([5, “Out of Limiter”] error occurs.) The limit value setting ranges from 1.0 to 550.0 Hz. Syntax: [SOURce:]FREQuency:LIMit:HIGH Parameters: 1.0 to 550.0 Command example: FREQ:LIM:HIGH 550 Query example: FREQ:LIM:HIGH? Response format: Response example: 550.
6.3 Command Descriptions [SOURce:]MODE This sets the output mode. There are eight output modes, each of them is a combination of one operation mode (AC or AC+DC) and one signal source mode, (INT, EXT, ADD, or SYNC). Consequently, the output mode setting is AC-INT, AC-EXT, AC-ADD, AC-SYNC, AC+DC-INT, AC+DC-EXT, AC+DC-ADD, or AC+DC-SYNC. However, when output is on, the output mode cannot be changed. ([1, “Invalid with output on”] error occurs.
6.3 Command Descriptions [SOURce:]PHASe[:IMMediate] This sets the phase at the start of output. The phase setting range is 0.0 to 359.9. However, when output is on, the phase setting cannot be changed. ([1, “Invalid with output on”] error occurs.) When in AC-EXT or AC+DC-EXT mode, the output waveform setting cannot be set. ([3, “Invalid in this mode”] error occurs.) Syntax: [SOURce:]PHASe[:IMMediate] Parameters: 0.0 to 359.
6.3 Command Descriptions [SOURce:]SEQuence:COUNt? This returns the jump times of the step specified by SEQ:STEP. The step jump times response ranges from 1 to 999. 1 is returned in the idle mode. Syntax: [SOURce:]SEQuence:COUNt? Parameters: None Query example: SEQ:COUN? Response format: Response example: 2 [SOURce:]SEQuence:DELete This clears the sequence memory. When this command is executed, all sequence step settings are cleared.
6.3 Command Descriptions Type Range 200.0 to 200.0 (400.0 to 400.0) Second parameter (DC voltage operation type) Type Range 0 (CONST), 1 (KEEP), or 2 (SWEEP) Third parameter (AC voltage [Vrms]/[Vp-p]) * The unit varies according to the waveform set by the seventh parameter. Type Range 0.0 to 140.0 [Vrms] or 0.0 to 400.0 [Vp-p] (0.0 to 280.0 [Vrms] or 0.0 to 800.
6.3 Command Descriptions [SOURce:]SEQuence:TPARameter This sets the step transition parameters. The target sequence step must be set in advance by the SEQ:STEP. ( See “[SOURce:]SEQuence:STEP” on page 6-22.) Use SYST:TUN ( See “ SYSTem:TUNit” on page 6-36) in advance to set the unit for the sequence step execution time or query the unit. When in AC-EXT, AC-SYNC, AC+DC-EXT, or AC+DC-SYNC mode, the step transition parameter cannot be set. ([3, “Invalid in this mode”] error occurs.
6.3 Command Descriptions [SOURce:]SEQuence:STEP Specifies the target step number for sequence editing. The step number specification ranges from 1 to 255. Syntax: [SOURce:]SEQuence:STEP Parameters: 1 to 255 Command example: SEQ:STEP 1 Query example: SEQ:STEP? Response format: Response example: 1 [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] AC output voltage is set using the Vrms unit (when the waveform is SIN or SQU) or the Vp-p unit (when the waveform is ARB).
6.3 Command Descriptions [SOURce:]VOLTage:LIMit:LOW This sets the upper limit of the output voltage that can be set. However, a value that is at or below the currently set cannot be specified. ([5, “Out of Limiter”] error occurs.) For the 100 V range, the limit value setting ranges from 200.0 to 5.0 V, and for the 200 V range, it ranges from 400.0 to 10.0 V. Syntax: [SOURce:]VOLTage:LIMit:LOW Parameters: 200.0 to 5.0 (400.0 to 10.
6.3 Command Descriptions d) MEASure subsystem MEASure[:SCALar]:CURRent[:AC]? This returns the output current RMS value (AC+DC RMS value) in Arms units. The output current RMS value measurement full scale is ±15.00 Arms. If the measured value exceeds the full scale, 99.99 Arms is returned as a measurement error. Syntax: MEASure[:SCALar]:CURRent[:AC]? Parameters: None Query example: MEAS:CURR? Response format: Response example: 12.
6.3 Command Descriptions MEASure[:SCALar]:CURRent:AMPLitude:MAX? This returns the output current peak value (hold) in Apk units. The held output current peak value full scale is 45.0 Apk. If the measured value exceeds the full scale, 99.9 Arms is returned as a measurement error. Syntax: MEASure[:SCALar]:CURRent:AMPLitude:MAX? Parameters: None Query example: MEAS:CURR:AMPL:MAX? Response format: Response example: 26.
6.3 Command Descriptions MEASure[:SCALar]:CURRent:HARMonic[:AMPLitude]? This returns the harmonic current in Arms units within the specified harmonic order range. The harmonic current measurement full scale is 15.00 Arms. If the measured value exceeds the full scale, 99.99 Arms is returned as a measurement error. However, when in AC-INT mode, if the frequency is neither 50 Hz nor 60 Hz, the harmonic current cannot be measured. ([3, “Invalid in this mode”] or [6, “Invalid in this frequency”] error occurs.
6.3 Command Descriptions MEASure[:SCALar]:POWer:AC:APParent? This returns the apparent power in VA units. The apparent power measurement full scale is 1400 VA. If the measured value exceeds the full scale, 9999 VA is returned as a measurement error. Syntax: MEASure[:SCALar]:POWer:AC:APParent? Parameters: None Query example: MEAS:POW:AC:APP? Response format: Response example: 367 MEASure[:SCALar]:POWer:AC:PFACtor? This returns the output power factor.
6.3 Command Descriptions MEASure[:SCALar]:VOLTage[:AC]? This returns the output voltage RMS value (AC+DC RMS value) in Vrms units. The output voltage RMS value measurement full scale is 450.0 Vrms. If the measured value exceeds the full scale, 999.9 Vrms is returned as a measurement error. Syntax: MEASure[:SCALar]:VOLTage[:AC]? Parameters: None Query example: MEAS:VOLT? Response format: Response example: 100.
6.3 Command Descriptions MEASure[:SCALar]:VOLTage:AVErage? This returns the output voltage average value (AC+DC average value) in V units. The output voltage average value measurement full scale is ±450.0 V. If the measured value exceeds the full scale, 999.9 V is returned as a measurement error. Syntax: MEASure[:SCALar]:VOLTage:AVErage? Parameters: None Query example: MEAS:VOLT:AVE? Response format: Response example: 0.
6.3 Command Descriptions f) STATus subsystem STATus:OPERation:CONDition? This returns the operation condition register (OPCR). The response ranges from 0 to 20482. Syntax: STATus:OPERation:CONDition? Parameters: None Query example: STAT:OPER:COND? Response format: Response example: 16384 STATus:OPERation:ENABle This sets the operation event enable register (OPEE). The setting ranges from 0 to 20482.
6.3 Command Descriptions STATus:OPERation:NTRansition STATus:OPERation:PTRansition This sets the operation transition filter (OPTF). The setting ranges from 0 to 20482. When the NTR filter is set to 1, OPER is set to 1 when OPCR changes from 1 to 0. When the PTR filter is set to 1, OPER is set to 1 when OPCR changes from 0 to 1. When the NTR and PTR filters are both set to 1, OPER is set to 1 when OPCR changes.
6.3 Command Descriptions STATus:WARNing:ENABle This sets the warning event enable register. The setting ranges from 0 to 31999. Syntax: STATus:WARNing:ENABle Parameters: 0 to 31999 Command example: STAT:WARN:ENAB 1 Query example: STAT:WARN:ENAB? Response format: Response example: 1 STATus:WARNing[:EVENt]? This returns the warning event register value. The response ranges from 0 to 31999.
6.3 Command Descriptions g) OUTPut subsystem OUTPut[:STATe] This controls on/off settings of the output relay. Syntax: OUTPut[:STATe] Parameters: OFF|0 / ON|1 Command example: OUTP ON Query example: OUTP? Response format: Response example: 1 h) INPut subsystem INPut:GAIN This sets the external input gain used in the external mode (AC-EXT or AC+DC-EXT) and internal + external mode (AC-ADD or AC+DC-ADD). For the 100 V range, the external input gain setting ranges from 0.
6.3 Command Descriptions i) TRACe subsystem TRACe:CATalog? This returns a list of defined waveform names. This command returns a list of all waveform names that can be selected by the user. Syntax: TRACe:CATalog? Parameters: None Query example: TRAC:CAT? Response format: [, ] Response example: SIN, SQU, ARB1, ARB2, ARB3, ARB4, ARB5, ARB6, ARB7, ARB8, ARB9, ARB10, ARB11, ARB12, ARB13, ARB14, ARB15, ARB16 TRACe[:DATA] This sends or receives arbitrary waveforms.
6.3 Command Descriptions TRACe:DELete[:NAME] This clears the specified arbitrary waveform memory. When this command is executed, a sine wave is written to ARB1 to ARB8, a square wave is written to ARB9 to ARB12, and a triangle wave is written to ARB13 to ARB16. However, when output is on, the arbitrary waveform memory cannot be cleared. ([1, “Invalid with output on”] error occurs.
6.3 Command Descriptions SYSTem:ERRor? This reads one error from the error queue (FIFO), and clears the read error in the error queue. The maximum number of errors that can be held in the error queue is 20. If 21 or more errors occur, [-350, “Too Many Errors”] is inserted as the 20th error. Syntax: SYSTem:ERRor? Parameters: None Query example: SYST:ERR? Response format: , Response example: 0, “No error” SYSTem:PON[:OUTPut] This sets the output status at power-on.
6.3 Command Descriptions SYSTem:WRELease This clears a warning. When a warning is cleared, the SET screen is displayed and key input is enabled. In order for the warning to be cleared, it is necessary that all the factors of the warning has been removed and it is ready to be cleared. (If no warning has occurred, or if a warning has occurred but the device is not ready to cancel the warning, [200, “Execution error”] error occurs.
6.3 Command Descriptions l) Common commands CLS This clears the following registers. Standard event register Operation event register Warning event register Status byte Error queue Syntax: CLS Parameters: None Command example: CLS Response format: None Response example: None ESE This sets the standard event enable register. The setting ranges from 0 to 255.
6.3 Command Descriptions IDN? This returns the model name, etc. The response message includes the “manufacturer, model name, serial number, and firmware version”. Syntax: IDN? Parameters: None Query example: IDN? Response format: Response example: “GW Instek , APS-1102, 000001, Ver1.00” OPC When all operations※ have been completed, this sets the standard event register OPC bit (BIT0). OPC? returns 1 when all operations have been completed.
6.3 Command Descriptions RST This resets the device to its factory settings. However, reset cannot be executed when output is on. ([1, “Invalid with output on”] error occurs.) Syntax: RST Parameters: None Command example: RST Response format: None Response example: None SAV This stores the current status information in memory. The memory storage ranges from 1 to 30.
6.3 Command Descriptions TST? This returns the self test results. In this device, “0” is always returned. Syntax: TST? Parameters: None Query example: TST? Response format: Response example: 0 WAI When all operations※ are completed, the device is not able to execute any subsequent commands. This is to prohibit the device from executing any subsequent command until all the operations are completed.
HIGH 6-42 LOW CONFigure EXTIO STATe CONDition PEAK BEEPer MODE MEASure [WINDow] DISPlay AVErage LIMit CURRent ERRor PON [EVENt] OPERation COUNt [IMMediate] OUTPut SYSTem ENABle CONDition CLOCk PHASe TUNit NTRansition CSTep EPARameter WRELease [AC] CONDition STATus PTRansition DELete SEQuence AVErage HIGH [AC] [EVENt] WARNing [DATA] TRACe STEP VOLTage ENABle CATalog LEN MODE [SOURce] MAX LOW AVErage [SCALar] MEASure HIGH GAIN INPut LOW HIGH HIGH
6.5 Status System 6.5 Status System The APS-1102 includes the minimum set of status groups defined by IEEE488.2. The APS-1102 status system is outlined below in Figure 6-4.
6.5 Status System 6.5.1 Status byte The status byte register definitions are listed in Table 6-13. Bits in the status byte register become valid when 1 is set in the service request enable register, then the ORed result of the valid bits is taken to issue a service request. A status byte can be read by the serial poll or an STB? query. Table 6-13.
6.5 Status System 6.5.2 Standard event status The structure of the standard event status register is shown in Figure 6-5 below. ESR (standard event register) Power on (PON) 7 7 User request (URQ) 6 6 Command error (CME) 5 5 Execution error (EXE) 4 4 Device specific error (DDE) 3 3 Query error (QYE) 2 2 Request control (RQC) 1 1 Operation complete (OPC) 0 0 Logical OR ESE (standard event enable register) Standard event status summary Status byte (bit 5) Figure 6-5.
6.5 Status System Table 6-14. Standard Event Status Register Definitions Bit Weight Description Power on bit PON (7) 128 URQ (6) 64 CME (5) 32 1 is set to this bit when the power is turned on. When this register is read, this bit is cleared to 0, and it remains 0 until the power is turned on again. User request bit Always 0 (not used) Command error 1 is set to this bit when a syntax error occurs in the program code.
6.5 Status System 6.5.3 Operation status The structure of the operation status register is shown in Figure 6-6 below.
6.5 Status System 6.5.4 Warning status The structure of the warning status register is shown in Figure 6-7 below.
6.6 Error Message List 6.6 Error Message List USB error messages are listed in Table 6-15 below. Table 6-15. USB Error Message List No. 0 Message “No error” Cause No error. Command was recognized as invalid due to any cause. 100 “Command error” 101 “Invalid character” Command was recognized as invalid due to invalid characters (such as “VOLT&”) in command string. 102 “Syntax error” Command was recognized as invalid due to reception of undefined command or parameter.
6.7 Programming Precautions 6.7 Programming Precautions a) Arbitrary waveform data Arbitrary waveform data comprises a block of binary data. The preceding ASCII program message command field and the binary data block arbitrary waveform data should be transferred at the same time. Always transfer arbitrary waveform data in 4096-word (8192-byte) batches. If the data count is insufficient, an error may occur, in which case the correct waveforms cannot be output.
7. T ROUBLESHOOTING 7.1 7.2 Protection Function ······························································ 7-2 Error Messages and Responses ··········································· 7-4 7.2.1 Error at power-on ·························································· 7-5 7.2.2 Protection function-related errors ································ 7-6 7.2.3 Panel operation errors ·················································· 7-8 7.2.
7.1 Protection Function 7.1 Protection Function The APS-1102 is equipped with a protection function that monitors the internal status and issues an error or warning when it detects the following types of abnormalities. If the beep sound has been set to on, a beep will accompany the displayed error or warning. (A beep sounds whenever a protection function-related error occurs, regardless of the beep sound setting.
7.1 Protection Function Table 7-1.
7.2 Error Messages and Responses 7.2 Error Messages and Responses The APS-1102 performs a self test at power-on, and outputs an error message if any abnormalities are detected. An error message is also displayed when a setting or operation is performed incorrectly or when an abnormality is detected in output or internally. The contents of these error messages, their causes, and the required responses are described below.
7.2 Error Messages and Responses 7.2.1 Error at power-on The APS-1102 performs the following fault checks at power-on. If an error occurs, NG is displayed on the system check screen. Nothing is displayed when all check items are OK. ROM CHECK...................... OK RAM READ/WRITE CHECK ........... OK BACKUP MEMORY CHECK ............ OK WAVE MEMORY CHECK .............. OK SEQUENCE MEMORY CHECK .......... OK CALIBRATION MEMORY CHECK ...... NG .... INITIALIZED VERSION CHECK .................. NG ....
7.2 Error Messages and Responses 7.2.2 Protection function-related errors When a protection function-related error has occurred, an error message is displayed in the message window, and operation is stopped. Figure 7-2. Screen Display When Protection Function-Related Error Has Occurred When a protection function-related error message is displayed, turn off the power switch. ( 7-6 See “7.
7.2 Error Messages and Responses Protection Function”.) Table 7-3 lists error messages and actions when errors are detected. Table 7-3. Protection Function-Related Errors Error Message System Locked. caused by: DCPS FAILURE System Locked. caused by: COMMUNICATION FAILURE 1 System Locked. caused by: COMMUNICATION FAILURE 2 System Locked. caused by: FRONT PANEL FAILURE System Locked.
7.2 Error Messages and Responses 7.2.3 Panel operation errors Panel operation error messages are displayed in the status window. The message box closes automatically after 1.5 seconds. Message box Figure 7-3. Screen Display When Panel Operation Error Has Occurred Table 7-4. Panel Operation Errors (1/2) Error Message OUT OF MEASURE RANGE INVALID IN THIS FREQUENCY OUT OF INPUT SIGNAL FREQUENCY OUT OF RANGE Cause and Action Required Measurement is out of range, cannot measure.
7.2 Error Messages and Responses Table 7-5. Panel Operation Errors (2/2) Error Message UNDER BUSY STATE Cause and Action Required Cannot set (execute) due to BUSY status. Wait until the BUSY display disappears, then try operation again. ( UNDER REMOTE STATE See “3.4.3 Setting output voltage range”.) Cannot set (execute) due to remote status. Set local mode. ( See “5.5.3 Remote”.) Cannot set (execute) due to local mode.
7.2 Error Messages and Responses 7.2.4 Warning messages When output is turned off due to an output error, internal temperature error, or external sync signal frequency out of range error, a warning is displayed in the message window. ( 7-10 See “7.
7.2 Error Messages and Responses Protection Function”.) Figure 7-4. Screen Display When Warning Has Occurred Until the warning has been canceled for the device, no key or command operations will be accepted. When the warning is canceled for the device, “Press ENTER key” is displayed on the warning screen. To close the warning screen, press the ENTER key or CANCEL key. If the “Press ENTER key.” message is not shown within 10 seconds, turn off the power and turn it on again.
7.2 Error Messages and Responses Table 7-7. Warning Messages (2/2) Message Output OFF. caused by: OVER PEAK CURRENT Output OFF. caused by: OVER DC VOLTAGE Output OFF. caused by: UNDER DC VOLTAGE Output OFF. caused by: REALTIME CONTROL FAILURE Output OFF. caused by: OVER TEMPERATURE OUTPUT OFF caused by: OUT OF SYNC FREQUENCY Cause and Action Required Output off due to the output peak current protection function. Press the ENTER key or CANCEL key. The error is cleared and the system becomes available.
7.3 When Fault Is Suspected 7.2.5 USB-related external control errors Table 7-6 lists USB-related external control error messages. Table 7-8. USB Error Message List No. 0 Message Text “No error” Cause No error. Command was recognized as invalid due to any cause. 100 “Command error” 101 “Invalid character” Command was recognized as invalid due to invalid characters (such as “VOLT&”) in command string.
7.3 When Fault Is Suspected Table 7-9. When Fault Is Suspected (Problem when Switching Power on/off) Symptom Possible Cause Required Action, etc. Not connected to a commercial Does not operate after power source. turning on power Power source being used is switch. outside the rated range. Internal fuse is broken. Does not stop immediately after turning off power switch. Make sure the power cord is firmly inserted into the outlet and the inlet (on rear panel of the APS-1102).
7.3 When Fault Is Suspected Table 7-11. When Fault Is Suspected (Problem Related to Output Voltage or Output Voltage Range Setting) Symptom Output voltage range cannot be switched. Possible Cause Required Action, etc. Turn power switch off. Fault in internal power source If the output voltage range cannot be switched (internal power source error has after turning the power source back on, contact occurred). Good Will or its representative for repairs.
7.3 When Fault Is Suspected Table 7-13. When Fault Is Suspected (Problem Related to Output Error) Symptom Possible Cause Required Action, etc. Output voltage setting is “0”. Signal source is not connected. External input gain is set to “0”. Check the output voltage setting. See “3.4.5 Setting output voltage”. When the signal source mode is external signal (EXT), connect the signal source to the external signal input terminal, then set a suitable external input gain. See “3.3.
7.3 When Fault Is Suspected Table 7-14. When Fault Is Suspected (Problem Related to Measurement Functions) Symptom Measured value for voltage or current is not displayed correctly. Possible Cause Required Action, etc. Selected display items are not appropriate. “----” is displayed. Measurement is out of range. (Out of measurable range) Measured values are inconsistent. Measurements are being taken in external mode (EXT). Measurements are at low frequency (10 Hz or less). For AC, select RMS.
7.3 When Fault Is Suspected Table 7-17. When Fault Is Suspected (Problem Related to Limiter Setting Range Limit) Symptom Output peak current limiter set value exceeds the output peak current limiter setting. Possible Cause Required Action, etc. The output peak current limiter may overshoot, Overshoot current value is being depending on the load. measured. See “3.4.7 Using limiter functions”. Table 7-18.
8. M AINTENANCE 8.1 8.2 8.3 8.4 8.
8.1 Introduction 8.1 Introduction This chapter describes the following items. Caution points and storage methods for long periods of non-use Caution points for transport and when repackaging for transport See “2.4 Simple Operation Checks”, for description of simple operation checks. If these operation checks are not passed, request calibration and/or repair from Good Will or its representative. 8.
8.2 Routine Maintenance b) When fan filter is dirty Two air filters are provided to filter out dust that gathers through the air inlets in the front panel. When too much dirt accumulates in the filters, it can clog them and reduce air flow, causing higher internal temperatures which can reduce reliability. Periodically check the condition of the air filters. The air filters should be cleaned about once a month, or whenever fine dust has accumulated in the filter.
8.3 Storage, Repackaging, and Transport 8.3 Storage, Repackaging, and Transport Store the APS-1102 in a location that meets the installation requirements. See “2.2 Installation Environment”, for description of installation requirements. a) Storage for long periods of non-use Remove the power cord from the outlet and the main unit. Store the main unit on a shelf or rack where it will be protected from airborne dirt and dust. When it may be subjected to dust, cover it with a cloth or polyethylene sheet.
8.4 Checking the Version Number 8.4 Checking the Version Number The APS-1102 firmware version is displayed in the bottom right of the screen for a few seconds after power-on. At other times, the version information can be checked via the control panel system information screen, as described below. Operation steps Select the SYSTEM menu from the MISC menu 1. Press the or CANCEL , then select “INFORMATION”. key to move the cursor to the icon. ENTER Cursor 2.
8.4 Checking the Version Number 4. Press the key to move the cursor to “INFORMATION”. CANCEL 5. Press the ENTER ENTER key to display the “SYSTEM INFORMATION” screen. CANCEL ENTER Figure 8-2. SYSTEM INFORMATION Screen 6. Press the TOP MENU key to return to the SET menu screen.
8.5 Backup Battery 8.5 Backup Battery The contents of memory ( See “4.8 Using Memory Functions”) and other required parameters and data are backed up by a lithium battery in the APS-1102. The backup period varies according to the temperature and use conditions, but it is approximately five years when stored with the power off. The backup data is checked at power-on, and an error message is displayed if the data has been corrupted ( See “7.
9. S PECIFICATIONS 9.1 9.2 9.3 9.4 9.5 9.6 9.7 9.8 9.9 9.10 9.11 9.12 9.13 9.14 9.15 9.16 9.17 9.
9.1 Output Unless otherwise noted, the following conditions are adopted: Power input: 200 V AC, 50 Hz Output mode: AC-INT (alternate current - internal signal source) mode Output frequency: 50 Hz Output voltage: 100 Vrms (200 Vrms when in 200 V range) Output waveform: Sine wave Load 100 V range: 10 Ω (power factor 1, nominal value) 200 V range: 40 Ω (power factor 1, nominal value) Output terminals: Terminal block on rear panel 9.
9.1 Output Output terminals Terminal block with M4 screws (rear panel) AC receptacle (universal type, front panel) Note: Use AC receptacle for AC (0 V AC to 250 V AC). When DC is included, use screw terminal block on rear panel.
9.1 Output Maximum output peak current 100 V range: 40 Apk 200 V range: 20 Apk Note 1: For the AC 100 V input system, the limit on the maximum output power may cause a reduction in the maximum output peak current. Note 2: The above values are for a capacitor input type rectifier circuit whose crest factor is 4 or less. Note 3: At or above the rated output voltage, the limit on the maximum output power may reduce the maximum output peak current.
9.1 Output Line regulation 0.
9.2 Current Limiter 9.2 Current Limiter Output peak current limiter When limiter is operating, output voltage is clipped. Positive current setting range 100 V range: +10.0 A to +42.0 A (Initial value: +42.0 A) 200 V range: +5.0 A to +21.0 A (Initial value: +21.0 A) 100 V range: 42.0 A to 10.0 A Negative current setting range: (Initial value: 42.0A) 200 V range: 21.0 A to 5.0 A (Initial value: 21.0 A) Setting resolution: 0.
9.4 Signal Sources 9.4 Signal Sources The signal sources that can be selected are internal (INT), external signal input (EXT), internal + external (ADD), and external synchronization (SYNC). Internal signal source (Internal source mode, internal + external source mode) See “9.1 Output”. External signal input (External signal source mode, internal + external source mode) Gain setting range 100 V range: 0.0 to 200.0 times (initial value: 100.0) 200 V range: 0.0 to 400.0 times (initial value: 200.
9.5 Measurement Functions 9.5 Measurement Functions Output voltage measurement RMS value: Full scale: RMS value of AC+DC 225.0 Vrms: 100 V range 450.0 Vrms: 200 V range Display resolution: 0.1 Vrms Measurement accuracy: ±0.5% of the full scale (at 23 ±5°C) Average value: Full scale: Average value of AC+DC (measures DC component) ±225.0 V: 100 V range ±450.0 V: 200 V range Display resolution: 0.1 V Measurement accuracy: ±0.
9.
9.6 Sequence Function 9.6 Sequence Function Output parameters can be changed abruptly or swept in sequence. Signal source is internal signal source only. Sequence function settings are stored in memory with battery backup. Number of sequences 1 sequence per operation mode (AC or AC+DC mode) and output range (100 V or 200 V) Number of steps 1 to 255 (within 1 sequence) Step time 0.1 ms to 999.9999 s (resolution 0.
9.7 Arbitrary Waveform Memory 9.7 Arbitrary Waveform Memory The contents of arbitrary waveform memory are retained by battery backup. Waveform memory count 16 Waveform length 4096 words Waveform data 15 bits Arbitrary waveform memory cannot be written to from the panel screen. USB interface is used for writing. 9.8 Setting Memory Various settings are retained in memory with battery backup, and the settings when the power source was last turned off are restored at power-on.
9.10 General 9.10 General LCD settings Contrast: Adjustable Display color: Blue background or white background Beep sound On or off When on, a beep sound will be output if a key input error or other operation error occurs. When a protection function-related error occurs, an alarm sound will be output regardless of the beep sound setting. Keylock On or off When keylock is on, only the keylock off operation and the output off operation are enabled.
9.11 External Control I/O 9.11 External Control I/O External control operation mode Enabled or disabled (status output is continually output) Control input Input level High level: +4.0 V or above, Low level: +1.
9.14 Withstand Voltage and Insulation Resistance 9.14 Withstand Voltage and Insulation Resistance Between input power line vs. output/case or input power line/case vs. output Withstand voltage 1500 V AC Insulation resistance 30 MΩ or above, 500 V DC 9.
9.17 External Dimensions and Weight %RH Storage condition 90 80 Performance guarantee 70 60 50 Operation guarantee 40 30 20 10 0 10 0 10 20 30 40 50 C Figure 9-1. Temperature and Humidity Ranges 9.17 External Dimensions and Weight External dimensions Width: 258 mm Height: 176 mm Depth: 440 mm (Projections are not included.) Weight Approximately 9.5 kg Figure 9-2 shows the external dimensions.
Built-in air filter Rear panel Protective cover 9-16 Surface treatment Control panel: Plastic ultra-light gray (Munsell 6. OPB9.2/0.1) Rear panel: Munsell 8.5PB2.6/0.2 Covers: Light gray leather tone (Munsell 6PB7.6/1.2 leather tone) Detailed view of cover labels Cover labels 9.17 External Dimensions and Weight Figure 9-2.
9.18 EC Declaration of Conformity 9.18 EC Declaration of Conformity We GOOD WILL INSTRUMENT CO., LTD. (1) No.7-1, Jhongsing Rd., Tucheng City, Taipei County, Taiwan (2) No.