Note: Changing the length will change the duty/
frequency of pulse waveforms.
4. Press F4 (N Cycle).
5. Press F1 (Cycles).
6. The Cycles property will become highlighted
in red.
7. Use the selector keys and
scroll wheel or number pad
to enter the number of
cycles.
/
8. Press F5 (Enter) to confirm
the number of cycles.
9. Press F6 (Return) to return
to the previous menu.
10. Press Trigger (F5) to
internally trigger the output
once.
Note: Ensure the output key has already been pressed
and the OUTPUT light is lit
before
pressing F5
(Trigger).
11. Press F6 (Return) to return
to the previous menu.
Below a pulse waveform of 5 cycles is output from the
front panel terminal.