User manual

AFG-200 SERIES USER MANUAL
Sync
A sync output is provided on the front panel SYNC port. All of the
standard output functions (except noise) have an associated Sync
signal. For Applications where you may not want to output the Sync
signal, you can disable the sync connector. The Sync signal may be
derived from either output channel in a two-channel instrument. By
default the Sync signal is derived from Channel 1.
OUTPut:SYNC
Instrument
Command
Description
Enables/Disables or queries the front panel Sync
output from the selected channel. The default is set
to off.
Syntax OUTPut:SYNC {OFF|ON}
Example OUTPut:SYNC ON
Turns the Sync output on.
Query Syntax OUTPut:SYNC?
Return Parameter
1 ON
0 OFF
Example OUTPut:SYNC?
1
The Sync output is on.
OUTPut[1|2]:SYNC:MODE
Instrument
Command
Description
This command is able to set the Sync output signal
in a different mode.
Syntax OUTPut[1|2]:SYNC:MODE {CARRier|MARKer}
Example OUTP1:SYNC:MODE CARR
Synchronizes the Sync output to the carrier signal
99