User manual

AFG-200 SERIES USER MANUAL
SOURce[1|2]:MARKer:FREQuency
Source Specific
Command
Description
Sets or queries the marker frequency. The default
marker frequency is 550 Hz. The marker frequency
is used to output a trigger out signal from the
trigger terminal on the front panel.
Note
The marker frequency must be between the start
and stop frequencies. If the marker frequency is set
to a value that is out of the range, the marker
frequency will be set to the center frequency and a
“settings conflict” error will be generated.
Syntax SOURce[1|2]:MARKer:FREQuency
{<frequency>|MINimum|MAXimum}
Parameter <frequency>
1μHz ~ 25 MHz
1 μHz ~ 1 MHz (Ramp)
Example SOUR1:MARK:FREQ +1.0000E+03
Sets the marker frequency to 1 kHz.
Query Syntax SOURce[1|2]:MARKer:FREQuency? [MINimum|
MAXimum]
Return Parameter
<NR3> Returns the marker
frequency in Hz.
Example SOUR1:MARK:FREQ? MAX
+1.0000E+03
Returns the marker frequency (1 kHz).