User manual
AFG-200 SERIES USER MANUAL
Example SOUR1:SWE:TIME?
+2.0000E+01
Returns the sweep time (20 seconds).
SOURce[1|2]:SWEep:SOURce
Source Specific
Command
Description
Sets or queries the trigger source as immediate
(internal)or manual. Immediate (internal) is the
default trigger source. IMMediate will constantly
output a swept waveform. Manual will ouput a
swept waveform after the trigger softkey is
pressed.
Note
If the APPLy command was used to create the
waveform shape, the source is automatically set to
IMMediate.
The *OPC/*OPC? command/query can be used to
signal the end of the sweep.
Syntax SOURce[1|2]: SWEep:SOURce {IMMediate|MANual}
Example SOUR1: SWE:SOUR IMM
Sets the sweep source to immediate.
Query Syntax SOURce[1|2]: SWEep:SOURce?
Return Parameter
IMM Immediate
MANual Manual
Example SOUR1:SWE:SOUR?
IMM
The sweep source is set to immediate.
76