User manual

AFG-200 SERIES USER MANUAL
Query Syntax SOURce[1|2]:SUM:INTernal:FUNction?
Return Parameter
SIN Sine UPRAMP Upramp
SQU Square DNRAMP Dnramp
TRI Triangle
Example SOUR1:SUM:INT:FUNC?
SIN
The shape for the modulating waveform is Sine.
SOURce[1|2]:SUM:INTernal:FREQuency
Source Specific
Command
Description
Sets the modulating waveform frequency for
internal sources. The default frequency is set to
100Hz.
Syntax SOURce[1|2]:SUM:INTernal:FREQuency
{<frequency>|MINimum|MAXimum}
Parameter <frequency> 2 mHz~ 20 kHz
Example SOUR1:SUM:INT:FREQ MAX
Sets the frequency to the maximum value.
Query Syntax SOURce[1|2]:SUM:INTernal:FREQuency?
Return Parameter
<NR3> Returns the frequency in
Hz.
Example SOUR1:SUM:INT:FREQ? MAX
+2.0000E+04
Returns the modulating frequency (20kHz).
SOURce[1|2]:SUM:AMPL
Source Specific
Command
Description
Sets or queries the amplitude of the modulating
waveform from the carrier waveform. The default
phase amplitude is 50%.
68