User manual

AFG-200 SERIES USER MANUAL
SOURce[1|2]:AM:STATe
Source Specific
Command
Description
Sets or disables AM modulation. By default AM
modulation is disabled. AM modulation must be
enabled before setting other parameters.
Note
Burst or sweep mode will be disabled if AM
modulation is enabled. As only one modulation is
allowed at any one time, other modulation modes
will be disabled when AM modulation is enabled.
Syntax SOURce[1|2]:AM:STATe {OFF|ON}
Example SOUR1:AM:STAT ON
Enables AM modulation.
Query Syntax SOURce[1|2]:AM:STATe?
Return Parameter
0 Disabled (OFF)
1 Enabled (ON)
Example SOUR1:AM:STAT?
1
AM modulation mode is currently enabled.
SOURce[1|2]:AM:INTernal:FUNCtion
Source Specific
Command
Description
Sets the shape of the modulating waveform from
sine, square, triangle, upramp and dnramp. The
default shape is sine.
Note
Square and triangle waveforms have a 50% duty
cycle. Upramp and dnramp have a symmetry of
100% and 0%, respectively.
Syntax SOURce[1|2]:AM:INTernal:FUNCtion
{SINusoid|SQUare|TRIangle|UPRamp|DNRamp}
Example SOUR1:AM:INT:FUNC SIN
Sets the AM modulating wave shape to sine.
51