User manual
AFG-200 SERIES USER MANUAL
impedance, the amplitude will double. Changing
the output termination from high impedance to
50Ω will half the amplitude.
If the output termination is set to high impedance,
dBm units cannot be used. The units will default to
Vpp.
Syntax OUTPut[1|2]:LOAD {DEFault|INFinity}
Example OUTP1:LOAD DEF
Sets the channel 1 output termination to 50Ω.
Query Syntax OUTPut[1|2]:LOAD?
Return Parameter DEF Default
INF INFinity
Example OUTP1:LOAD?
DEF
The output termination for channel 1 is set to 50Ω.
SOURce[1|2]:VOLTage:UNIT
Source Specific
Command
Description
Sets or queries the output amplitude units. There
are three types of units: VPP, VRMS and DBM.
Note
The units set with the VOLTage:UNIT command
will be used as the default unit for all amplitude
units unless a different unit is specifically used for
a command.
If the output termination is set to high impedance,
dBm units cannot be used. The Units will
automatically default to Vpp.
Syntax SOURce[1|2]:VOLTage:UNIT {VPP|VRMS|DBM}
Example SOUR1:VOLT:UNIT VPP
Sets the amplitude units to Vpp.
Query Syntax SOURce[1|2]:VOLTage:UNIT?
Return Parameter VPP Vpp
46