AFG-200 SERIES USER MANUAL
Query Syntax OUTPut[1|2]:SYNC:MODE?
Return Parameter CARR
The output is
synchronized with the
carrier.
MARK
For scan, pulse, and
arbitrary waveforms the
marker signal is output
from the Sync output.
Example OUTP1:SYNC:MODE?
CARR
Indicates that the Sync output signal is synchronized
to the carrier signal.
OUTPut[1|2]:SYNC:POLarity
Instrument
Command
Description
This command controls the Sync signal output
polarity.
Note
The normal Sync signal polarity is from a TTL low
to high level. The inverted Sync signal polarity is
from a TTL high to low level.
Syntax OUTPut[1|2]:SYNC:POLarity{INVerted|NORMal}
Example OUTP1:SYNC:POLarity INV
Sets the Sync output signal polarity to inverted.
Query Syntax OUTPut[1|2]:SYNC:POL?
INV
Inverted
Return Parameter
NORM
Normal
Example OUTP1:SYNC:POL?
INV
Indicates that the output Sync signal for CH1 is
inverted.
100