User manual

AFG-2000 Series User Manual
124
SOURce[1]:SWEep:STATe
Set
Query
Sets or disables Sweep mode. By default sweep is
disabled. Sweep must be enabled before setting
other parameters.
Any modes will be disabled if sweep mode is
enabled.
SOURce[1]:SWEep:STATe {OFF|ON}
SOUR1:SWE:STAT ON
Enables sweep mode.
SOURce[1]:SWEep:STATe?
0
Disabled (OFF)
1
Enabled (ON)
SOUR1:SWE:STAT?
>1
Sweep mode is currently enabled.
SOURce[1]:FREQuency:STARt
Set
Query
Sets the start frequency of the sweep. 100Hz is the
default start frequency.
To sweep up set the stop frequency higher than the
start frequency. Set the stop frequency lower than
the start frequency to sweep down.
SOURce[1]:FREQuency:STARt
{<frequency>|MINimum|MAXimum}
<frequency>
0.1Hz ~ 25MHz*
0.1Hz ~ 1MHz (Ramp)
*AFG-2112 limited to 12MHz, AFG-2105 limited to
5MHz.