User manual

REMOTE INTERFACE
103
FSK
SWEEP
SOURce[1]:FUNCtion {SINusoid|SQUare|RAMP
|NOISe |USER}
SOUR1:FUNC SIN
Sets the output as a sine function.
SOURce[1]:FUNCtion?
SIN, SQU, RAMP, NOIS,
USER
Returns the current output
type.
SOUR1:FUNC?
>SIN
Current output is sine.
SOURce[1]:FREQuency
Set
Query
Sets the output frequency for the
SOURce[1]:FUNCtion command. The query
command returns the current frequency setting.
The maximum and minimum frequency depends
on the function mode.
Sine, Square
0.1Hz~25MHz*
Ramp
0.1Hz ~ 1MHz
Noise
Not applicable
User
0.1Hz~10MHz*
*AFG-2005/2105 limited to 5MHz, AFG-2012/2112
limited to 12MHz.