Datasheet

10
Good Will Instrument Co., Ltd.Simply Reliable
Turbo OFF, OCP Istep 3.75 A Istop 37.5A
Test result screen
Turbo OFF, OCP Istep 3.75 A Istop 37.5A
Setting
Turbo OFF, OCP Istep 3.75 A Istop 37.5A
The actual test waveform
Turbo ON, OCP Istep 7.5 A Istop 75.0A
The actual test waveform
Turbo ON, OCP Istep 7.5 A Istop 75A
Test result screen
Turbo ON, OCP Istep 7.5 A Istop 75.0A
Setting
Turbo : ON, Fuse ON, CC pulse 75A, 1S,
Test 3 cycles the actual test waveform
Turbo : OFF, Fuse ON, CC pulse 37.5A, 2S,
Test 3 cycles the actual test waveform
TurboOFF, Fuse mode
Test result screen
Turbo ON, Fuse mode
Test result screen
SettingTurboOFF, Fuse ON
CC pulse 37.5A, 2S, Test 3 cycles
SettingTurboON, Fuse ON
CC pulse 75.0A, 1S, Test 3 cycles
Basically, Fuse test has Trip (Blown) and Non-Trip (no Blown) 2 types.
Fuse Test setting parameters include test current (Istart), test time (Time), test REPEAT Time etc..
In the Trip fuse test, it is used to test when there is too large abnormal current the Fuse or Bleaker must be able to provide the protection
of the circuit break, that means current protection components need the fuse action, therefore the test current needs to be larger than
the fuse current rating.
When the AEL-5000 Series AC & DC electronic load detects a voltage lower than 1.0V, the LCD displays the number of Repeat Cycle and
Current Protection Fusing Time XXXX.X sec.
In the Non-Trip (no Blown) test, the current protection component is required to achieve non-blow action, so the test current needs to
be lower than the fuse current rating that is used to verify the fuse must not blow during normal current range. When the AEL-5000
Series AC & DC electronic load is not blown after the test time (Pulse Time) and the repeated Repeat number, the LCD displays the
information of the Repeat number.
AC & DC Electronic Load