Manual

OPERATION
65
The load can be ―on‖ when switching between
A Value and B Value.
Display:
Static Mode
Mode
CP
I Range
H 35A
V Range
H 150V
Function
Static
Configure
0.00
V
0.000
A
0.00
w
CP B Value 0.00
CP A Value 0.00
LOADRS232
A Value
Fine
03/Sep/2012
W
W
Active setting
A Value
B Value
+CV OFF V
Static mode
Display:
Dynamic Mode
Mode
CP
I Range
H 35A
V Range
L 15V
Function
Dynamic
Configure
0.000
V
0.000
A
0.00
w
Timer1 0.025
Level2 0.00
Level1 0.00
LOADRS232
Fine
03/Sep/2012
ms
W
W
Dynamic mode
Level1
Level2
Timer1
Select the Display Units for Dynamic Mode Levels
Description
When Dynamic switching mode is selected, the
Level1 and Level2 values can be set to either
discrete values or as a percentage of a set value.
The setting applies to all applicable operation
modes.
By default the units are set to Value.
When Percent is chosen, 100% = 100% of the Set
power, current or resistance value.