Manual
EXTERNAL CONTROL
137
CR Mode
Input conductance =
rated conductance × (external voltage/10)
Input
Conductance
External
Voltage
Rated
Conductance
0V 10V
CV Mode
Input voltage = rated voltage × (external
voltage/10)
Input
Voltage
External
Voltage
Rated
Voltage
0V 10V
CP Mode
Input power = rated power × (external
voltage/10)
Input Power
External
Voltage
Rated
Power
0V 10V
Operation
1. Turn the power off from the PEL-3000 and from
the load.
2. Connect the external voltage across pins 1 and 3
of the J1 connector.