Manual
PEL-3000 User Manual
110
Fast Sequence Overview
Description
A fast sequence is comprised of a user-defined
number of steps that can be executed at a high
frequency. Unlike normal sequences, each step
in a fast sequence has the same execution time
(time base).
This mode is only available for CC and CR
mode.
Up to 1000 discrete steps can be configured
using fast sequences.
Each fast sequence can have a memo note
attached to it.
Fast Sequences can be looped up to 9999 discrete
times or for an infinite amount of times.
Fast sequences can be configured to hold a set
current or resistance at the end of the load.
No ramping function can be used with the Fast
Sequence function.
Fast Sequence
STEP 01
STEP 02
:
STEP N
Loop: