    COMMAND DETAILS 
177 
Example 
:SEARCH:TRIGger:RISEFall:WHEn UNEQual 
Sets the rise andfall search trigger condition to 
unequal (≠). 
:SEARCH:TRIGger:RISEFall:TIMe 
Set
Query
Description 
Sets or queries the rise and fall time. 
Syntax 
:SEARCH:TRIGger:RISEFall:TIMe {<NRf> | ? } 
Related 
commands 
:SEARCH:TRIGger:TYPe 
:SEARCH:TRIGger:RISEFall:WHEn 
Parameter 
<NRf> 
Rise and Fall time (4nS to 10S) 
Return Parameter 
<NR3> 
Returns the rise and fall time in 
seconds. 
Example 
:SEARCH:TRIGger:RISEFall:TIMe 4.00E-5 
Sets the trigger rise and fall time to 40.0us. 
:SEARCH:TRIGger:LOGic:INPut:CLOCK:SO
URce 
Set
Query
Description 
Sets or queries which channel is used as the clock 
source for the logic search trigger. 
Note 
When “NONe” is selected as the clock source, the 
trigger will use the Pattern search trigger type. 
Syntax 
:SEARCH:TRIGger:LOGic:INPut:CLOCK:SOURce 
{NONe | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D 
9 | D10 | D11 | D12 | D13 | D14 | D15 | ? }  
Related 
commands 
:SEARCH:TRIGger:LOGic:INPut:CLOCK:SOURce 
:SEARCH:TRIGger:LOGic:INPut:CLOCK:EDGe 
Parameter/Return 
parameter 
NONe 
No clock source, Set to Pattern search 
trigger. 
D0 to D15 
Digital channels D0 to D15 