User's Manual
1514
1.When the monitor is off,press the
START/STOP button to turn on the monitor,
and it will finish the whole measurement,
save and transmit the measurement data for
the desired user. (Take User 1 for example.)
Adjust the zero .
LCD display
Inflating and measuring.
Display and save the results. The
data transmission will proceed.
Start the Measurement
4.Press “START/STOP” button to
power off, otherwise it will turn
off within 1 minute.
MEASUREMENT
Tips:
A. You can press “START/STOP” button at any time to stop measuring during the
process of measurement.
B. Maximum 60 records are both for USER 1 and USER 2.
C. If the measurement result is out of the measurement range (SYS: 60mmHg to
230mmHg; or DIA: 40mmHg to 130mmHg;or Pulse: 40-199 pulse/minute), the
LCD will display “out”.
2.This device will proceed
to data transmission
after measurement.
The Bluetooth symbol
blinks on the LCD
indicates data is
transmitting.
3.If the data is successfully
transmitted, the Bluetooth
symbol doesn’t blink any more,
and the will disappear.
If the data transmission fails,
the will proceed.
START
STOP
START
STOP
CAUTION
1.
2.
Interference may occur in the vicinity of equipment marked with the following symbol
. And TMB-995-BS may interfering vicinity electrical equipment.
Sensitive people, including pregnant women pre-eclamptic and those who implanted
medical electronic instruments, should avoid using the unit whenever possible.
Keep the monitor at least 20 centimeters away from the human body (especially the
head) when the data transmission is proceeding after measurement.
To enable the data transmission function, this product should be paired to Bluetooth
end at 2.4 GHz.
How to mitigate possible interference?
The range between the device and BT end should be reasonably close, from 1 meter
to 10 meters. Please ensure no obstacles between the device and BT end so as to
obtain quality connection and to lower the RF output range.
To avoid interference, other electronic devices (particularly those with wireless
transmission / Transmitter) should be kept at least 1 meter away from the monitor.
MEASUREMENT