User's Manual
Guidance and manufacturer’s declaration – electromagnetic Immunity
Immunity Test
±8 kV contact
±2 kV, ±4kV, ±8 kV, ±15 kV air
±8 kV contact
±2 kV for power supply lines
±0.5 kV, ±1 kV differential mode
Table 2
Compliance level
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC61000-4-5
±1 kV signal input/output
100 kHz repetition frequency
IEC 60601-1-2
Test level
±2 kV, ±4kV, ±8 kV,
±15 kV air
±2 kV for power supply lines
Not Applicable
100 kHz repetition frequency
±0.5 kV, ±1 kV,±2 kV common
mode
±0.5 kV, ±1 kV differential mode
Not Applicable
27