Version:1.0 User Manual Wrist Blood Pressure Monitor TMB-1598-BS Wrist Type Guangdong Transtek Medical Electronics Co., Ltd. Zone A, No.105, Dongli Road, Torch Development District, 528437 Zhongshan, Guangdong, China Thank you very much for selecting TRANSTEK Wrist Blood Pressure Monitor TMB-1598-BS. To use the monitor correctly and safely, please read the manual thoroughly. Please keep this manual well in order to reference in future.
CATALOGUE INTRODUCTION ............................................................................................. 2 General Description Indications for Use Contraindications Measurement Principle Safety Information LCD Display Symbol Monitor Components List BEFORE YOU START ......................................................................................
INTRODUCTION INTRODUCTION General Description Thank you for selecting TRANSTEK Blood Pressure Monitor (TMB-1598-BS). The monitor features blood pressure measurement, pulse rate measurement and the result storage. The design provides you with two years of reliable service. Readings taken by the TMB-1598-BS are equivalent to those obtained by a trained observer using the cuff and stethoscope auscultation method.
INTRODUCTION INTRODUCTION Safety Information The signs below might be in the user manual, labeling or other component. They are the requirement of standard and using. Refer to instruction manual/booklet To signify that the instruction manual/booklet must be read.
INTRODUCTION INTRODUCTION CAUTION 6 *On the rare occasion of a fault causing the cuff to remain fully inflated during measurement, open the cuff immediately. Prolonged high pressure (cuff pressure > 300mmHg or constant pressure > 15mmHg for more than 3 minutes) applied to the wrist may lead to an ecchymosis. *Please check that operation of the device does not result in prolonged impairment of patient blood circulation. * When measurement, please avoid compression or restriction of the connection tubing.
INTRODUCTION CAUTION 8 * Cleaning :Dust environment may affect the performance of the unit. Please use the soft cloth to clean the whole unit before and after use. Don’t use any abrasive or volatile cleaners. * The device doesn’t need to be calibrated within two years of reliable service. * If you have any problems with this device, such as setting up, maintaining or using, please contact the SERVICE PERSONNEL of Transtek. Don’t open or repair the device by yourself in the event of malfunctions.
INTRODUCTION INTRODUCTION SYMBOL DESCRIPTION kPa mmHg Irregular heartbeat Current time 10 EXPLANATION Measurement Measurementunit unitofofthe theblood bloodpressure pressure Monitor Components (1kPa=7.5mmHg) Measurement unit the blood pressure (1mmHg=0.133kPa) Blood pressure monitor is detecting an irregular heartbeat during measurement.
INTRODUCTION Component list of pressure measuring system: 1. PCBA; 2. Air Pipe; 3. Pump; 4. Valve; 5. Cuff. BEFORE YOU START Installing and Replacing the Batteries CUFF (Type BF applied part) • Slide off the battery cover. • Install the batteries by matching the correct polarity, as shown below. Always use the correct battery type (2×AAA batteries.) • Replace the cover. The typical service life of the new and unused batteries is 54 measurements for the operation time is 60s.
BEFORE YOU START BEFORE YOU START Replace the batteries whenever the below happens The shows The display is dim. The display does not light up Setting Date, Time, Positioning Symbol and Measurement Unit It is important to set the clock before using your blood pressure monitor, so that a time stamp can be assigned to each record that is stored in the memory. (The setting range of year is 2016 ~ 2056. Time format: 24H) 1. When the monitor is off, press “ ” button, it will display the time.
BEFORE YOU START Notes: 1.During the process of setting, you can press “ ” button to stop setting at any time. 2. If there is no operation during the process of setting, it will turn off within 1 minute. BEFORE YOU START 4.Repeat step 2 and 3 to confirm [MONTH] and [DAY]. 2. Press “ ” button to change the [YEAR]. Each press will increase the numeral by one in a cycling manner. 3 . When you get the right year, press “ ” button to confirm and it willINFORMATION turn to next step. FOR USER 16 5.
BEFORE YOU START 6.Repeat step 2 and 3 to confirm [Positioning symbol] on or off. BEFORE YOU START BEFORE YOU START 8.After confirming the meausrement unit, the LCD will display “donE”, then display all the settings you have done one by one and then the monitor will turn off. 7.Repeat step 2 and 3 to confirm the measurement unit.
BEFORE YOU START Select the User ID BEFORE YOU START MEASUREMENT 2. After confirming the user ID, the LCD will display “User ID+donE” and then turn off. Before you start the measurement, please select the desired user ID first. 1. When the blood pressure monitor is off, press and hold “ ” button until the user ID blinks. Then press “ ” button to change the user ID between user 1 and user 2. Press “ SET ” button to confirm your selection.
BEFORE YOU START BEFORE YOU START Pair up the Blood Pressure Monitor with Your Device 1.Turn on Bluetooth and the app. Make sure both are ON when pair-up is proceeding. ” button to 2 .When the monitor is OFF, press and hold the “ start pair-up. The symbol “ ” and the symbol “ ” will be shown on the LCD alternindicating pair-up is proceeding. If SUCCEED, the symbol “ be shown on the LCD. 22 + ” will If FAIL, the symbol “ shown on the LCD. + ” will be 3.
MEASUREMENT Tie the Cuff 1. Remove all accessories (watch, bracelet,etc) from your wrist. If your physician has diagnosed you with poor circulation in your wrist, use the other one. 2. Roll or push up your sleeve to expose the skin. 3. Apply the cuff to your wrist with your palm facing up. 4. Position the edge of the cuff about 1cm~1.5cm from wrist joints. 5. Fasten the wrist cuff around your wrist, leaving no extra room between the cuff and your skin.
MEASUREMENT 2.After finishing the measurement, the LCD displays the blinking icon ‘ ’,and transmits the data. Note: (1). The symbol “ ” will disappear after successful data transmission, and you may check your personal health data stored in your mobile device. (2). If the data transmission fails, the symbol “ ” will remain. The pending measurement data will be transmitted to your mobile device when next measurement is complete. (3).
DATA MANAGEMENT DATA MANAGEMENT DATA MANAGEMENT Recall the Records 3. If you want to check the other user’s measurement records, please press ” button to turn off the blood pressure “ monitor. Then press and hold “ ” button to enter the selecting user mode, press “ ” again to change the user, when the desired user ID blinks, press “ ” button to confirm. Then press“ ” button to check the records of the selected user. 1.
DATA MANAGEMENT Delete the Records If you did not get the correct measurement, you can delete all the results for the selected user by following steps. (Take user 1 for example.) DATA MANAGEMENT 3. If there is no record, press “ ” button he below display will be shown. 1. Hold pressing “ ” button about 3 seconds when the monitor is in the memory recall mode , the “dEL All” +User ID will flash on the display. 2. Press “ ” button to confirm ,the LCD displays “ User ID+dEL donE” and then turns off.
INFORMATION FOR USER INFORMATION FOR USER Maintenance Tips for Measurement To obtain the best performance, please follow the instructions below. Measurements may be inaccurate if taken in the following circumstances. Within 1 hour after dinner or drinking Immediate measurement after tea, coffee, smoking Put in a dry place and avoid the sunshine Within 20 minutes after taking a bath Avoid immersing it in the water. Clean it with a dry cloth in case. Avoid shaking and collision.
EMC GUIDANCE ABOUT BLOOD PRESSURE ABOUT BLOOD PRESSURE What are systolic pressure and diastolic pressure? Irregular Heartbeat Detector What is the standard blood pressure classification? An irregular heartbeat is detected when a heartbeat rhythm varies while the device is measuring systolic pressure and diastolic pressure. During each measurement, blood pressure monitor will keep a record of all the pulse intervals and calculate the average value of them.
ABOUT BLOOD PRESSURE TROUBLESHOOTING Why does my blood pressure fluctuate throughout the day? 1. Individual blood pressure varies multiple times everyday. It is also affected by the way you tie your cuff and your measurement position, so please take the measurement under the same conditions. 2.If the person takes medicine, the pressure will vary more. PROBLEM SYMPTOM 3.Wait at least 3 minutes for another measurement.
TROUBLESHOOTING PROBLEM SYMPTOM Error message EExx,shows on the display. Warning message out shows SPECIFICATIONS CHECK THIS A calibration error occurred. (XX can be some digital symbol,such as 01,02,etc., if this similarsituation appear, all belong to calibration error.) Out of measurement range REMEDY Retake the measurement.If the problem persists,contact the retailer or our customer service department for further assistance. Refer to the warranty for contact information and return instructions.
SPECIFICATIONS CONTACT INFORMATION Contact Information Mode of operation Continuous operation Degree of protection Type BF applied part Device Classification Internally Powered ME Equipment IP Classification Software Version IP22: The first number 2: Protected against solid foreign objects of 12,5mm Ф and greater. The second number: Protected against vertically falling water drops when enclosure titled up to 15º.
EMC GUIDANCE EMC Guidance The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare environments Table 1 Guidance and manufacturer’s declaration - electromagnetic emissions Warning:Don’t near active HF surgical equipment and the RF shielded room of an ME system for magnetic resonance imaging, where the intensity of EM disturbances is high. Emissions test Warning: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation.
EMC TableGUIDANCE Table 55 EMC EMC GUIDANCE GUIDANCE Table 2 Guidance and manufacturer’s declaration – electromagnetic Immunity Immunity Test IEC 60601-1-2 Test level Compliance level Electrostatic discharge (ESD) IEC 61000-4-2 ±8 kV contact ±2 kV, ±4kV, ±8 kV, ±15 kV air ±8 kV contact Electrical fast transient/burst IEC 61000-4-4 Not applicable Not applicable Surge IEC61000-4-5 ±2 kV, ±4kV, ±8 kV, ±15 kV air Voltage dips, short interruptions and voltage variations on power supply input lines I
TableGUIDANCE 5 EMC EMC GUIDANCE Table 3 Guidance and manufacturer’s declaration - electromagnetic Immunity Radiated RF Test IEC61000-4-3 Frequency (MHz) (Test specifications 385 for ENCLOSURE PORT IMMUNITY to 450 RF wireless communications 710 equipment) 745 780 810 870 930 46 Band (MHz) Service Modulation 380-390 TETRA 400 Pulse modulation 18 Hz 430-470 GMRS 460, FM ± 5k Hz deviation FRS 460 1 kHz sine 704-787 LTE Band 13, 17 Pulse modulation 217 Hz Pulse 800-960 GSM 800/900, modulation TETRA 8