User's Manual
This section includes a list of error messages and
frequently asked questions for problems you may
encounter with your blood pressure monitor. If the
products not operating as you think it should, check here
before arranging for servicing.
Power supply
Battery powered mode:
6VDC 4*AA alkaline batteries
AC adaptor powered mode:6V 1A
(Can be supplied by AC adaptor model
UE08WCP-060100SPA !)
Display mode
Digital LCD V.A.68mm*90mm
Measurement mode
Oscillographic testing mode
Measurement range
Accuracy
Normal working condition
Storage & transportation
condition
Measurement perimeter
of the upper arm
About 22cm~42cm
Net Weight
Approx.300g(Excluding the dry cells)
External dimensions
Attachment
Approx.92mm*140mm*46mm
4*AA alkaline batteries,user manual,AC adapter
Mode of operation
Continuous operation
Degree of protection
Software Version
Type BF applied part
Protection against
ingress of water
V01
WARNING: No modification of this equipment is allowed.
IP22
Rated cuff pressure:
0mmHg~299mmHg(0kPa ~ 39.9kPa)
Measurement pressure:
SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa)
DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa)
Pulse value: (40-199)beat/minute
Pressure:
5℃-40℃ within±3mmHg
pulse value:±5%
Temperature:-20°C to +60°C
A relative humidity range of ≤ 93%, non-condensing,
at a water vapour pressure up to 50hPa
PROBLEM SYMPTOM CHECK THIS REMEDY
No power
Low
batteries
Error
message
Display
will not light up.
Batteries are exhausted.
Replace with new batteries
Insert the batteries
correctly
Replace with new batteries
Batteries are inserted
incorrectly.
Batteries are low.
shows
Unsuccessful pairing.
Check if both the APP and
Bluetooth are on, operate
and send the data again.
E 1 shows
The cuff is not secure.
Readjust the cuff and
relax for a moment
and then measure again.
AC adaptor is inserted
incorrectly.
Insert the AC adaptor
tightly
The display
indicates the
“BAT LO”
message, pauses
for 3 seconds.
The battery icon
shows empty
(does not flash.)
A temperature range of :+5°C to +40°C
A relative humidity range of 15% to 90%,
non-condensing, but not requiring a water
vapour partial pressure greater than 50 hPa
An atmospheric pressure range of :
700 hPa to 1060 hPa
Warning
message
“out ” shows
Out of measurement
range
Relax for a moment.
Refasten the cuff and then
measure again. If the
problem persists, contact
your physician.
TROUBLESHOOTING
SPECIFICATIONS
1918