version:1.0 User Manual Blood Pressure Monitor TMB-1585-BS Arm Type FCC ID:OU9TMB1585B2 Guangdong Transtek Medical Electronics Co., Ltd. Zone A, No.105 ,Dongli Road, Torch Development District, 528437 Zhongshan,Guangdong,China Thank you very much for selecting TRANSTEK Blood Pressure Monitor TMB-1585-BS. Please read the user manual carefully and thoroughtly so as to ensure the safe usage of this product, Keep the manual well for further reference in case you have problems.
CATALOGUE Table of Contents INTRODUCTION...................................................................................................................2 General Description Indications for Use Contraindications Measurement Principle Safety Information LCD Display Signal Monitor Components List BEFORE YOU START...........................................................................................................
INTRODUCTION INTRODUCTION General Description Thank you for selecting TRANSTEK arm type blood pressure monitor (TMB-1585-BS). The monitor features blood pressure measurement, pulse rate measurement and the result storage. The design provides you with two years of reliable service. Readings taken by the TMB-1585-BS are equivalent to those obtained by a trained observer using the cuff and stethoscope auscultation method.
INTRODUCTION CAUTION * This device is intended for indoor, home use. * This device is not intended for public use. * This device is portable, but it is not intended for use during patient transport. * This device is not suitable for continuous monitoring during medical emergencies or operations. * This device is intended for no-invasive measuring and monitoring of arterial blood pressure.
INTRODUCTION INTRODUCTION LCD display signal Monitor Components CUFF Component list of pressure measuring system 1 Cuff 2 Air pipe 3 PCBA 4 Pump 5 Valve AIR HOSE AIR CONNECTOR PLUG LCD DISPLAY SYMBOL DESCRIPTION Systolic blood pressure High blood pressure Diastolic blood pressure Low blood pressure Pulse display Pulse in beats per minute Indicate it is in the memory mode and which group of memory it is. Motion may result in an inaccurate measurement.
BEFORE YOU START BEFORE YOU START The Choice of Power Supply Setting Date, Time and Measurement Unit 1.Battery powered mode: It is important to set the clock before using your blood pressure monitor, so that a time stamp can be assigned to each record that is stored in the memory. (The setting range of the year :2017—2057 time format:24H) 6VDC 4×AAA batteries 2.AC adaptor powered mode: 6V 1A (Please only use the recommended AC adaptor model).
BEFORE YOU START BEFORE YOU START 4.Repeat step 2 and 3 to set the [MONTH] and [DAY] . 6 .Repeat step 2 and 3 to set the [UNIT]. 5 .Repeat step 2 and 3 to set the [HOUR] and [MINUTE]. 7 .After the unit is set,the LCD will display “donE” first, then display all the settings you have done and then turn off.
BEFORE YOU START Select the User 1.When the monitor is off , press and hold “M” button to enter user setting mode. The user ID will blink. MEASUREMENT Tie the Cuff 1. Remove all jewelry, such as watches and bracelets from your left arm.Note: If your doctor has diagnosed you with poor circulation in your left arm, use your right arm. 2. Roll or push up your sleeve to expose the skin. Make sure your sleeve is not too tight. 3.
MEASUREMENT MEASUREMENT Start the Measurement 2.This device will proceed 1.When the monitor is off, press the “START/STOP” button to turn on the monitor, and it will finish the whole measurement. (Take user 1 for example.) LCD display Adjust the zero. to data transmission after measurement. The Bluetooth symbol blinks on the LCD indicates data is transmitting. If the data transmission fails, the LCD will display . Inflating and measuring. If the data transmission succeeds, the with disappear.
DATA MANAGEMENT DATA MANAGEMENT 3. If you want to check another user’s records, press “START/STOP” button to turn off the monitor when the blood pressure monitor is in the memory recall mode. Press and hold “M” button to enter into the selecting user ID mode, press “M” again to select the user ID between user 1 and user 2 press “S” button to confirm the user ID , then press “M” button to check the selected user’s measurement records. Recall the Records 1.
INFORMATION FOR USER DATA MANAGEMENT Delete the Records Tips for Measurement If you did not get the correct measurement, you can delete all results by following steps below. (Take User 1 for example.) 1.Hold pressing “M” button for 3 seconds when the monitor is in the memory recall mode ,the flash display “ User ID+ dEL ALL” will show. Note: To exit out the delete mode without deleting any records, press START/STOP button before pressing "S" button to confirm any delete commands.
INFORMATION FOR USER ABOUT BLOOD PRESSURE What are systolic pressure and diastolic pressure? Maintenance In order to get the best performance, please follow the instructions below. When ventricles contract and pump blood out of the heart, the blood pressure reaches its maximum value in the cycle, which is called systolic pressure. When the ventricles relax, the blood pressure reaches its minimum value in the cycle, which is called diastolic pressure.
ABOUT BLOOD PRESSURE TROUBLESHOOTING Why does my blood pressure fluctuate throughout the day? 1. Individual blood pressure varies multiple times everyday. It is also affected by the way you tie your cuff and your measurement position, so please take the measurement under the same conditions. 2.If the person takes medicine, the pressure will vary more. 3.Wait at least 3 minutes for another measurement.
AUTHORIZED COMPONENT SPECIFICATIONS Battery powered mode: 6VDC 4×AAA batteries Power supply AC adaptor powered mode: 6V 1A (Please only use the recommended AC adaptor model). Digital LCD V.A.100mm×68mm Display mode Oscillographic testing mode Measurement mode Rated cuff pressure: 0mmHg~299mmHg(0kPa ~ 39.9kPa) Measurement pressure: Measurement range SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa) DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa) Pulse value: (40-199)beat/minute Pressure: 5°C-40°C within±3mmHg(0.
EMC GUIDANCE FCC STATEMENT EMC Guidance FCC Statement FCC ID:OU9TMB1585B2 This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
EMC GUIDANCE EMC GUIDANCE Table 2 Guidance and manufacturer’s declaration – electromagnetic Immunity Immunity Test IEC 60601-1-2 Test level Electrostatic discharge (ESD) IEC 61000-4-2 ±8 kV contact ±2 kV, ±4 kV, ±8 kV, ±15 kV air ±8 kV contact ±2 kV, ±4 kV, ±8 kV, ±15 kV air Electrical fast transient/burst IEC 61000-4-4 ±2 kV for power supply lines ±1 kV signal input/output 100 kHz repetition frequency ±2 kV for power supply lines Not Applicable 100 kHz repetition frequency Surge IEC61000-4-5 ±0.