User's Manual

13
12
MEASUREMENT
BEFORE YOU START
BEFORE YOU START
CAUTION
1.
2.
Interference may occur in the vicinity of equipment marked with the following symbol .
And TMB-1583-BS may interfering vicinity electrical equipment.
Sensitive people, including pregnant women pre-eclamptic and those who implanted
medical electronic instruments, should avoid using the unit whenever possible.
Keep the monitor at least 20 centimeters away from the human body (especially the
head) when the data transmission is proceeding after measurement.
To enable the data transmission function, this product should be paired to Bluetooth end
at 2.4 GHz.
How to mitigate possible interference?
The range between the device and BT end should be reasonably close, from 1 meter to
10 meters. Please ensure no obstacles between the device and BT end so as to obtain
quality connection and to lower the RF output range.
To avoid interference, other electronic devices (particularly those with wireless
transmission / Transmitter) should be kept at least 1 meter away from the monitor.
Pair-up the Blood Pressure Monitor
with Your Device
1.Turn on Bluetooth and the app. Make sure both are ON when pair-up is proceeding.
2
If SUCCEED,
the symbol + ”
and user ID will be shown on the LCD.
If FAIL,
the symbol + ”
and user ID will be shown on the LCD.
3.The monitor will shut off after Pair-up process is complete.
.When the monitor is OFF, press and hold the START/STOP button to
start pair-up. The symbol “ + ”and user ID will be shown on the LCD
alternatively, indicating pair-up is proceeding.
Bluetooth Module No.: LS51802
RF Frequency Range: 2402 MHz to 2480 MHz
Output Power Range: -2.6 dBm
Supply Voltage:
2V-3.6 V
Transmitting Distance: 10 meters
List of compatible devices:
For iOS devices:
The operating system must be iOS 8 or more, such as iPhone
4S, iPhone 5/5C/5S, iPhone 6/6 Plus and so on.
For Android devices:
The operating system must be 4.3 or more.