User's Manual

MEM
SET
START
STOP
In order to get the best effect and protect your monitor,please use the right
battery and special power adaptor which complies with U.S. safety standard.
.Open the battery cover.
.Install the batteries by matching
the correct polarity, as shown.
.Replace the cover.
Installing and Replacing the Batteries
CAUTION
Remove batteries if the device is not likely to be used for some time.
The old batteries are harmful to the environment, do not dispose with other daily trash.
Remove the old batteries from the device and follow your local recycling guidelines.
CAUTION
Replace the batteries whenever the below happen
The shows
The display dims
The display does not light up
Do not dispose of batteries in fire. Batteries may explode or leak.
AC adaptor
The Choice of Power Supply
1.Battery powered mode:
6VDC 4×AAA alkaline batteries
2.AC adaptor powered mode:
6V 1A
(Please only use the recommended AC
adaptor model).
Please unplug the adaptor to depart from
the using utility power.
Measurement Principle
It is important to set the clock before using your blood pressure
monitor, so that a time stamp can be assigned to each record that
is stored in the memory. (The setting range of the year :2014—2054
time format:12H/24H)
1.When the monitor is off,
hold pressing “SET” button
for 3 seconds to enter the
mode for year setting.
2.Press the “MEM” button to
change the [YEAR]. Each
press will increase the
numeral by one in a
cycling manner.
This product uses the Oscillometric Measuring method to detect blood pressure. Before
every measurement, the unit establishes a “zero pressure” equivalent to the air pressure.
Then it starts inflating the arm cuff, meanwhile, the unit detects pressure oscillations
generated by beat-to-beat pulsatile, which is used to determine the systolic and diastolic
pressure, and also pulse rate.
The device also compares the longest and the shortest time intervals of detected pulse
waves to mean time interval then calculates standard deviation. The device will display a
warning signal with the reading to indicate the detection of irregular heartbeat when the
difference of the time intervals is over 25%.
Setting Date, Time, Measurement Unit
and Clock Mode
BEFORE YOU START
BEFORE YOU START
9