Version:1.0 User Manual Blood Pressure Monitor TMB-1580-BT Wrist Type Guangdong Transtek Medical Electronics Co., Ltd. Zone A, No.105 ,Dongli Road, Torch Development District, Zhongshan,528437,Guangdong,China Thank you very much for selecting TRANSTEK Blood Pressure Monitor TMB-1580-BT. To use the monitor correctly and safely, please read the manual thoroughly. Please keep this manual well in order to reference in future.
CATALOGUE CATALOGUE INTRODUCTION ............................................................... 2 General Description Indications for Use Safety Information LCD Display Signal Monitor Components BEFORE YOU START .......................................................... 5 Installing and Replacing the Batteries Setting Date, Time and Measurement Unit Pair-up the Blood Pressure Monitor with Your Device MEASUREMENT ...............................................................
INTRODUCTION INTRODUCTION General Description CAUTION Thank you for selecting TRANSTEK Blood Pressure Monitor (TMB-1580-BT). The monitor features blood pressure measurement, pulse rate measurement and the result storage. The design provides you with two years of reliable service. Readings taken by the TMB-1580-BT are equivalent to those obtained by a trained observer using the cuff and stethoscope auscultation method.
INTRODUCTION INTRODUCTION Monitor Components LCD Display Signal SYSTOLIC LCD DISPLAY GRADE TIME MEM BUTTON SYMBOL DESCRIPTION Systolic blood pressure High pressure result Diastolic blood pressure Low pressure result Pulse Pulse/minute; heartbeats/minute Shocking remainding Shocking will result in inaccurate Low battery Batteries are low and need to be replaced kPa mmHg 4 EXPLANATION Measurement unit of the blood pressure (1kPa=7.5mmHg) Measurement unit the blood pressure (1mmHg=0.
BEFORE YOU START BEFORE YOU START Installing and Replacing the Batteries • .Slide off the battery cover. • .Install the batteries by matching the correct polarity, as shown below. Always use the correct battery type (2 x AAA batteries). • .Replace the cover. Setting Date, Time and Measurement Unit It is important to set the clock before using your blood pressure monitor, so that a time stamp can be assigned to each record that is stored in the memory. (year :2014—2054,time format:24 H/12 H) 1.
BEFORE YOU START 4.Repeat steps 2 and 3 to confirm [HOUR] and [MINUTE]. 5.Then the monitor diverts to date format setting. Select the date format. Repeat steps 2 and 3 to confirm the date format between DM and MD. BEFORE YOU START 7.Repeat steps 2 and 3 to confirm the [MEASUREMENT UNIT]. 8.After confirming the meausrement unit, the LCD will display “dOnE” and then turn off. 6.Repeat steps 2 and 3 to confirm the [MONTH] , [DAY] and [YEAR].
BEFORE YOU START BEFORE YOU START Pair-up the Blood Pressure Monitor with Your Device 1 .Turn on Bluetooth and the app. Make sure both are ON when pair-up is proceeding. 2 .When the monitor is OFF, press and hold START/STOP button about 3 seconds to start 4 .The monitor will shut off automatically after Pair-up process is complete. Bluetooth Module No.: AW2540MV1 RF Frequency Range: 2402 MHz to 2480 MHz Output Power Range: -1 dBm pair-up. The symbol Supply Voltage: 2 V to 3.
MEASUREMENT Tie the Cuff 1 .Remove all accessories (watch, bracelet, etc) from your wrist. If your physician has diagnosed you with poor circulation in your wrist, use the other one. 2.Roll or push up your sleeve to expose the skin. 3.Apply the cuff to your wrist with your palm facing up. 4. Position the edge of the cuff about 1-1.5cm. 5. Fasten the wrist cuff around your wrist, leaving no extra room between the cuff and your skin. If the cuff is too loose, the measurement will not be accurate.
MEASUREMENT DATA MANAGEMENT Start the Measurement 1.When the monitor is off, press “START/STOP” button to turn on the monitor, and it will finish the whole measurement. (Take User 1 for example.) LCD display Inflating and measuring. 2. This device will proceed to data transmission after measurement. The Bluetooth symbol blinks on the LCD indicates data is transmitting. Adjust to zero. Display and save the result. If the data transmission fails, lights all the time.
DATA MANAGEMENT DATA MANAGEMENT Delete the Records Recall the Records If you did not get the correct measurement, you can delete all results by following steps below . (Take User 1 for example.) 1. When the monitor is off, press “M” button to show the average value of the latest three measurement records. (Take User 1 for example.) 1. In the memory mode, hold pressing “M” button for 3 seconds, the flash display “User ID+ dEL All” will show. 2.
INFORMATION FOR USER INFORMATION FOR USER Maintenance Tips for Measurement Measurements may be inaccurate if taken in the following circumstances. Wait at least 1 hour after dinner or drinking Immediate measurement after tea, coffee, smoking In order to get the best performance, please follow the instructions below.
ABOUT BLOOD PRESSURE ABOUT BLOOD PRESSURE What are systolic pressure and diastolic pressure? When ventricles contract and pump blood out of the heart, blood pressure reaches its maximum value, the highest pressure in the cycle is known as systolic pressure. When the heart relaxes between heartbeats, the lowest blood pressure is diastolic pressure.
TROUBLESHOOTING SPECIFICATIONS This section includes a list of error messages and frequently asked questions for problems you may encounter with your wrist blood pressure monitor. If the products not operating as you think it should, check here before arranging for servicing. PROBLEM SYMPTOM No power Low batteries Display is dim or will not light up. REMEDY Batteries are exhausted. Replace with new batteries Batteries are inserted incorrectly. Show on Batteries are low.
CONTACT INFORMATION Contact Information For more information about our products, please visit www.transtek.cn.you can get customer service, usual problems and customer download, transtek will serve you anytime. Manufactured by: Guangdong Transtek Medical Electronics Co., Ltd. Company: Guangdong Transtek Medical Electronics Co., Ltd. Address: Zone A, No.
COMPLIED STANDARDS LIST EMC GUIDANCE Complied Standards List Risk management ISO/EN 14971:2012 Medical devices — Application of risk management to medical devices Labeling ISO/EN 15223-1:2012 Medical devices. Symbols to be used with medical device labels, labelling and information to be supplied.