26
Table 3
Guidance and manufacturer’s declaration - electromagnetic Immunity
Radiated RF
IEC61000-4-3
(Test
specifications
for ENCLOSURE
PORT
IMMUNITY to
RF wireless
communications
equipment)
Test
frequency
(MHz)
Band
(MHz)
Service Modulation
Maximum
Power (W)
Distance
(m)
Test
Level
(V/m)
Compliance
level
(V/m)
385 380-390 TETRA 400
Pulse
modulation
18 Hz
1,8 0.3 27 27
450 430-470
GMRS 460,
FRS 460
FM ± 5 kHz
deviation
1 kHz sine
2 0.3 28 28
710
704-787
LTE Band
13, 17
Pulse
modulation
217 Hz
0.2 0.3 9 9745
780
810
900-960
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
18 Hz
2 0.3 28 28
870
930