Use Instructions

25
Table 2
Guidance and manufacturer’s declaration -
electromagnetic Immunity
Immunity Test IEC 60601-1-2
Test level
Compliance level
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8
kV, ±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±8
kV, ±15 kV air
Electrical fast
transient/burst
IEC 61000-4-4
Not applicable Not applicable
Surge
IEC 61000-4-5
Not applicable Not applicable
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Not applicable Not applicable
Immunity Test IEC 60601-1-2
Test level
Compliance level
Power
frequency
magnetic field
IEC 61000-4-8
30 A/m
50Hz/60Hz
30 A/m
50Hz/60Hz
Conducted RF
IEC61000-4-6
Not applicable Not applicable
Radiated RF
IEC61000-4-3
10 V/m
80 MHz – 2,7
GHz
80 % AM at 1
kHz
10 V/m
80 MHz – 2,7
GHz
80 % AM at 1
kHz
NOTE UT is the a.c. mians voltage prior to application of
the test level.