User's Manual

1514
Table 1 Guidance and manufacture’s declaration – electromagnetic emissions-
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration – electromagnetic emission
Emission test Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
Class B
Not applicable
Not applicable
Compliance
The LS802-E is intended for use in the electromagnetic environment specified
below. The customer of the user of the LS802-E should assure that it is used in
such and environment.
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions IEC
61000-3-3
RF emission
CISPR 11
The LS802-E is intended for use in the electromagnetic environment specified
below. The customer or the user of LS802-E should assure that it is used in such an
environment.
Guidance and manufacture’s declaration – electromagnetic immunity
Immunity test
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
±2 kV for
power supply lines
±2kV for
power supply lines
±1 kV line(s)
to line(s)
±1 kV
differential mode
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
)
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
3A/m
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
Table 2 Guidance and manufactures declaration – electromagnetic immunity –
for all ME EQUIPMENT and ME SYSTEMS
IEC 60601 test level Compliance level
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile. If
floor are covered with
synthetic material, the
relative humidity should be
at least
30%.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the LS802-E requires
continued operation during
power mains interruptions,
it is recommended that the
LS802-E be powered from
an uninterruptible power
supply or a batter
y.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
3A/m
Power frequency
(50Hz) magnetic
field IEC
61000-4-8
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines IEC
61000-4-11
Electrical fast
transient/burst
IEC 61000-4-4
Surge IEC
61000-4-5
The
LS802-E
uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
The LS802-E is suitable for use in all
establishments other than domestic and
those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes.