User's Manual

1716
Table 4 Guidance and manufacture’s declaration – electromagnetic immunity –
for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration – electromagnetic immunity
Immunity test
3 Vrms
3 V/m
The LS802-E
is intended for use in the electromagnetic environment specified below.
The customer or the user should assure that it is used in such an environment.
Compliance
level
IEC 60601
test level
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V/m 80 MHz
to 2.5 GHz
3 V
rms
150 kHz to
80 MHz
Electromagnetic environment -
guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the
YS-6100
, including
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Portable and mobile RF communications
equipment should be used no closer to
any part of the including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
d = 1.167
d = 1.167
80 MHz to 800 MHz
d
= 2.333
800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended
separation distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should be
less than the compliance level in each
frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.
a Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoreticall
y with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in
the location in which the ELE007839V1 is used exceeds the applicable RF
compliance level above, the ELE007839V1 should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating
the ELE007839V1.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.