Operator's Manual

ETD-100 OPERATOR’S MANUAL
53
10.2 Basic Troubleshooting
Basic troubleshooting can be completed without the need for a trained service technician. The User can complete
several tasks to ensure the device is functioning as intended. Errors that are known to occasionally occur include:
Blank Swabs resulting in false positives following a threat sample test due to contamination remaining after a
WARNING run
Initialization is taking longer than expected
Sample tests are inconclusive several times in succession
Verification Pen tests are returning incorrect results
The Operator can follow the following troubleshooting steps when an error occurs. If the error persists after the
Operator completes each of the tasks, a service technician may be required.
1. Run a manual flush by pressing the flush button on the Settings Screen
2. Replace the Sample Filter
3. Replace the Sample Trap Holder
4. Switch off the device. Once the device is turned off, turn on again and allow full initialization process
5. If the problem persists, call the GreyScan service team
10.3 Error Messages
If an error occurs, the User will be presented with an Error Screen:
Figure 63: Example error Screen
The error code, details about the error and a suggested remedy is displayed. Below is a list of all the possible errors.
Code
Detail
Remedy
C11
Clamp fault
Call service technician