Operator's Manual

ETD-100 OPERATOR’S MANUAL
33
7 Storage & Transport
7.1 Device Preparation for Storage and Transport
To prepare the device for short term storage or transport:
1. Power off the device by pressing the POWER button.
2. Disconnect the AC Power Cable from the wall power outlet.
3. Disconnect the DC Power Cable from the device DC Power Port by rotating the locking clasp anti-clockwise
and removing the DC Power Cable.
4. Remove the Sample Trap Holder and Sample Filter.
5. Cover the Sample Filter and Sample Trap Holder slots with Sellotape or Kapton tape.
6. Remove the Reagent Cartridge. Refer to Section 8.4 Changing the Reagent Cartridge.
7. Install the transport protection caps over each fluidic port. These caps were removed during installation and
setup. Ensure the red cap is installed onto the top port with the red marking. Refer to Section 5.2 Installation
of the Reagent Cartridge.
Note: Steps 4, 5 and 6 are not required if the device is simply being moved between locations for use at a different
place.
8. If the Pelican Case has been supplied with a mesh storage area, then verify all loose components are
securely stowed.
9. Close the device Enclosure Lid and lock the clips in place. To lock the clips, push them down and towards the
device Enclosure. They will click into place when secure.
Figure 33: Device Enclosure Top View (Closed)