Operator's Manual
ETD-100 OPERATOR’S MANUAL
26
Alert thresholds are factory configurable and are documented in the device configuration manual supplied
with the device.
Highlight Color
Corresponding Detection Level
Red
High Level (≥100% of configured threshold)
Orange
Medium Level (33% - 99% of configured threshold)
Yellow
Low Level (5% - 32% of configured threshold)
Grey
No Detection (0% - 5% of configured threshold)
Blue
Internal Standard
Table 2: Electropherogram Highlight Color Details
Examples of colored Electropherograms can be seen in Figure 23 and Figure 24.
Figure 23: Electropherogram with Yellow and Blue Highlight
Figure 24: Electropherogram with Red and Blue Highlights
6.3.3 REMOVE SWAB Screen
The REMOVE SWAB Screen will display after a sample test
WARNING and the User has proceeded from the WARNING
Screen. This screen will prompt the User to remove the Swab
and discard.
Once the User has removed the Swab, the PROCEED button
is used to begin the full device flush to prepare
for a new test sample.
Figure 25: Remove Swab Screen