5 Installation and Setup
5.1 Setup Method
1. Unpack the device and Reagent Cartridge
• Ensure the cardboard packaging and support material are retained for future shipment.
2. Install the Reagent Cartridge (If not already fitted)
• Refer to Section 5.2 Installation of the Reagent Cartridge.
3. Install the Sample Trap Holder (If not already fitted)
• Refer to Section 5.3 Installation of the Sample Trap Holder.
4. Install the Sample Filter (If not already fitted)
• Refer to Section 5.4 Installation of the Sample Filter and Sample Filter Retainer
5. Plug in the power cable
• Plug the DC power adapter plug into the device DC power port.
• Insert the AC plug of the power adapter into a wall power outlet.
• The Meanwell GST90A24-P1M power adapter and cable suitable must not be substituted with any other
power cable or adapter.
6. Turn on the device
• Press the power button on the device. It should illuminate red.
• The PIN code Screen will take approximately 30 seconds to load.
• Enter the password PIN code provided with the GreyScan ETD-100 Quick Start Guide.
7. Calibration
• Once the device is turned on, allow the device to complete internal calibration/initialisation tests. This will take
10-15 minutes to complete.
• When internal calibration is complete, the INSERT SWAB Screen will be displayed, and the device is ready
for use.