Installation Instructions GGO-case:
a.) Dismantling of sensors
- unscrew cable buckling protection and push it upwards the
cable.
(loosen the cable in the remaining buckling protection part at
the rubber sealing gasket).
- unscrew both case parts.
- push case part at cable side upwards the cable and disconnect the sensor connector.
- unscrew the sensor carefully e.g. by means of suitable nippers.
b.) Assembly of new sensors
- remove new sensor element from packing.
- check exact position of sealing O-ring on the thread
- screw the sensor element carefully into the corresponding half of the case and tighten slightly e.g. by means of
nippers.
Attention: do not use force! Case can be destroyed by this!
- reconnect sensor plug to sensor element.
- screw on the second half of the case and the buckling protection.
Installation Instructions GOO-case:
a.) Dismantling of sensors
- unscrew cable buckling protection and push it upwards the
cable.
(loosen the cable in the remaining buckling protection part at
the rubber sealing gasket).
- unscrew lock screws if existing.
- unscrew both case parts.
- push case part at cable side upwards the cable and disconnect the sensor connector.
- unscrew the sensor-element carefully.
b.) Assembly of new sensors
- remove new sensor element from packing.
- check exact position of sealing ring on the thread
- screw the sensor element carefully into the corresponding half of the case and tighten slightly.
- reconnect sensor plug to sensor element.
- screw on the second half of the case and lock screws (if existing).
- screw on the buckling protection.
Installation Instructions GOX 100 sensor case:
- disconnect the sensor connector
- open screw cap, memorize the positions of the rubber rings
- remove the sensor from case
- remove new sensor element from packing.
- remove O-ring from new sensor (not needed)
- Place new sensor in case, considering the rubber rings
- Close screw cap, reconnect
Disposal of sensor element:
The sensor contains lead and caustic electrolyte. Dispose as special waste.
According to the ElektroG (law for bringing into market, the return and the environmentally friendly disposal
of electronic equipment) we accept the return of this sensors, please send it directly to us (adequately
stamped). We will dispose it appropriately and environmentally friendly.