User manual

H60.0.12.6C-13 Operating Manual GMH 3851 Page 16 of 22
_____________________________________________________ _____________________________________________________________________________
13 Specification
Measuring Channel1 Channel2
Principle Resistive material-moisture-measuring Temperature-measuring thermocouple type K
matching DIN EN 13183-2: 2002 or internal temperature-measuring
Char. curve 466 different kinds of wood matching DIN EN 60584-1: 1996, ITS90
28 different building materials
4 individually programmable material curves
Probe connection BNC Plug floating connector for mini-blade-terminal
Meas. range 0.0...100.0 % moisture content thermocouple: -40.0... +200.0°C / -40.0... + 392.0°F
(depending on characteristic curve) int. temp.-Meas.: -30.0...75.0°C / -22.0...167.0°F
equal to ca. 3kOhm ... 2TerraOhm
Resolution 0,1% moisture content 0.1°C / 0,1°F
Evaluation Evaluation of the material condition in 9 steps from DRY to WET
Accuracy Device without probe ±1Digit (at nominal-temperature)
Wood: ±0.2% moisture content (deviation from Type K: ± 0.5% m.v. ± 0.3°C
characteristic curve, range 6..30%) int. t.-measuring: ± 0.3°C (is type K reference junction)
building mat.: : ±0.2% moisture content (dev. from
char. curve, range depending on char. curve)
Temperature drift < 0.005 % moisture content per 1K 0.01% per 1K
Nominal temperature 25°C
Ambient Temperature -25 ... +50°C (-13 .. 122°F)
Relative humidity 0 ... 95 %RH (non condensing)
Storage temperature -25 ... +70°C (-13 ... 158°F)
Housing Dimension: 142 x 71 x 26 mm (L x B x D)
impact resistant ABS, membrane keyboard, transparent panel.
Front side IP65, integrated pop-up-clip for table top or suspended use
Weight approx. 155 g
Output: 3.5mm audio plug, stereo
Selectable as serial interface: via optically isolated interface adapter GRS3100, GRS3105 or USB3100 (p.r.t. accessories)
directly connectable to RS232- or USB-interfaces.
or analogue output: 0..1V, freely scalable (resolution 13bit, accuracy 0.05% at nominal temperature, cap. load <1nF)
Real time clock: Integrated clock with date and year
Logger: 2 Functions: individual value logger („Func–Stor“) and cyclic logger („Func–CYCL“)
Memory: Stor: 99 data sets; CYCL: 10000 data sets
Cycle time CYCL: 0:30...60:00 (minutes:seconds, min 1s, max 1h)
Power Supply 9V-Battery, type IEC 6F22 (included) as well as additional d.c. connector (diameter of internal pin 1.9 mm) for
external 10.5-12V direct voltage supply. (suitable power supply: GNG10/3000)
Power Consumption output off approx. 2.5mA
output serial interface: approx. 2.7mA
analogue output: approx. 3.0mA
cyclic logger sleeping state with output deactivated: < 0.1mA
cyclic logger sleeping state with activated serial interface: < 0.3mA
Display Two 4 digits LCD’s (12.4mm high and 7 mm high) for material moisture temperature or characteristic curve,
hold function, etc. as well as additional pointing arrows.
Pushbuttons 6 membrane keys for on/off switch, menu operation, characteristic curve, hold-function etc.
Hold Function Press button to store current value.
Automatic-Off-Function Device will be automatically switched off if no key is pressed/no interface communication takes place for
the time of the power-off delay. The power-off delay can be set to values between 1 and 120 min.; it can be
completely deactivated.
EMC: The device corresponds to the essential protection ratings established in the Directives of the European
Parliament and of the council on the approximation of the laws of the member states relating to the
electromagnetic compatibility (2004/108/EC).
EN61326 +A1 +A2 (Appendix B, class B), additional error: < 1% FS
14 Disposal Notes
Dispense exhausted batteries at destined gathering places.
This device must not be disposed as “residual waste”. To dispose this device, please send it
directly to us (adequately stamped). We will dispose it appropriately and environmentally friendly.