User manual

H60.0.71.6C-04 Manual Appendix GMH 38-LW1 / -LW2 page 5 of 8
_____________________________________________________ _____________________________________________________________________________
9 Measuring of grain
For the storability, their quality assessment and purpose, the 38-LW Set is an important support of decision
beside the decision of smell (fusty?) Consistence (dust…) and appearance (colour, impurities).
When measuring grain pls keep in mind to use sufficient amount of measuring good (at least 500ml), sur-
rounding the sensor and that there is sufficient pressure between sensor and grain (in heap >30 cm and an
immersion depth of >20 cm this is usually automatically the case), otherwise there may be to low display
values!
With freshly harvested grain, an approximate recommendation for barley, rye und wheat can be given:
Less than 16 % u Measured material is sufficiently dry and storable.
16 - 20 % u Measured material includes increased humidity, appropriate dry before storage.
Above 20 % u extreme high humidity! Defer harvesting if possible or dry before storage
10 Wood Chips as fuel
Instrument settings for measuring wood chips:
GMH 3830/3850/3851 Version>= 1.5: h.461 (specialised GSF 38 / GSF 50 curve)
others: We recommend “Wood group C” (GMH 38x0 instruments: “h. C”). This group delivers a sufficient
accuracy for the fuel application up to 30% MC above there is larger deviation.
Wood chips are classified in different quality groups.
The size and the moisture content (MC or u) or the wet-basis moisture content (w) is the measure for the
usability. Usually moisture content (w) of maximum 30% is recommended.
Chip size
Class
Size
G 30
small chips
smaller than 3 cm
G 50
mid size chips
3 5 cm
G 100
crude chips
5 10 cm
Moisture content
Class
Wet basis moisture content w
(can be displayed directly of GMH 3830 V>=1.5)
moisture content u
w 20
air dry
<20 % w
<25 % u
w 30
storable
20 30 % w
25 43 % u
w 35
conditionally storable
30 35 % w
43 54 % u
w 40
wet
35 40 % w
54 67 % u
w 50
fresh cut
40 50 % w
67 100 % u
The higher the moisture content, the lower is the heating value per weight
10.1 Field measuring
At measuring in containers, silos, chip bunkers or similar storages and a measuring depth > 0.5 m commonly
the compression is high enough for direct measuring.
Although keep pressure on the handle during the measuring!
For measuring in less than 0.5 m or in loose bulk material, best is to step on the measuring spot and insert
the probe below the foot.
At values above 20%u the display may have falling values: The display after 10 seconds is valid!