User manual
H32.0.0X.6C-06  Operating Manual GMSD … - K31  page 4 of 4 
6  Specification:  GMSD … .. - K31 
6.1  Specification (GMSD … BA - K31): 
    GMSD 1,3 BA - K31  GMSD 2 BA - K31  GMSD 7 BA - K31 
Measuring range:    0 ... 1300 mbar abs.  0 ... 2000 mbar abs.  0,00 ... 7,00 bar abs. 
Overload:    max. 4 bar abs.  max. 4 bar abs.  max. 10 bar abs.   
Resolution:    1 mbar  1 mbar  0,01 bar  (10 mbar) 
Accuracy: (typ. values)  ±0,2 %FS (hysteresis and linearity) 
±0,4%FS  (temperature influence from 0-50 °C) 
OPTION:
 higher accuracy
:  ±0.1 %FS (hyst., linearity); ±0.2%FS (temperature influence 0-50 °C) 
Sensor:  piezoresistive absolute pressure sensor. For air pressure (barometer), vacuum, absolute 
pressure. Suitable for air and/or non-corrosive and non-ionising gas. 
6.2  Specification (GMSD … ..R - K31): 
GMSD
2,5
MR
-
K31  GMSD
25
MR
-
K31 GMSD
350
MR
-
K31  GMSD
2
BR
-
K31  GMSD
10
BR
-
K31
Measuring range:    -1,999 ... 2,500 mbar  -19,99 ... 25,00 mbar  -199,9 ... 350,0 mbar  -1000 ... +2000 mbar  -1,00 ... +10,00 bar 
  (-199,9 ... 250,0 Pa)  (-1999 ... 2500 Pa) 
Overload:    max. 200 mbar  max. 300 mbar  max. 1 bar  max. 4 bar  max. 10,34 bar 
Resolution:    0,001 mbar  0,01 mbar  0,1 mbar  1 mbar  0,01 bar 
  (0,1 Pa)  (1 Pa) 
Accuracy: (typ. values)    (0-2,5mbar)  (0-25mbar)  (0-350mbar)  (0-2bar)  (0-10bar) 
 Hysteresis and linearity:    ±0,2%FS  ±0,2%FS  ±0,2%FS  ±0,2%FS  ±0,2%FS 
Temperature influence from 0-50°C
:  ±1,0%FS  ±0,5%FS  ±0,4%FS  ±0,4%FS  ±0,4%FS 
OPTION:
 higher accuracy
:        ±0,1%/±0,2%FS  ±0,1%/±0,2%FS  ±0,1%/±0,2%FS 
Sensor:  piezoresistive relative pressure sensor. For over-/under- and difference pressure measuring. 
Suitable for air and/or non-corrosive and non-ionising gas. 
If sensor is to be used in water, use air cushion! 
6.3  Common specifications (GMSD … BA - K31, GMSD … ..R - K31): 
Pressure connection:  2 nylon connecting pins for tubes 6 x 1 mm (6 mm outer Ø, 4mm internal Ø) 
Electronics:  PC-board with amplifier and data memory for sensor data (measuring data, calibration etc.) 
integrated in sensor housing. 
Nominal temperature:  25 °C   
Operating conditions:  0 ... +50 °C,  0 ... +95 %RH (non condensing) 
Storage temperature:  -25 ... +70 °C 
Housing:  68 x 32,5 x 15 mm (L x W X D) without connection pin;  
68 x 32,5 x 27,5 mm incl. connection pin. 
ABS housing with integrated suspension eye 
Device Connection:  1m PVC connection cable, screened with integral 6-pin MiniDIN plug, lockable. 
Weight:  approx. 75 g   
EMC:  The GMSD ... corresponds to the essential protection ratings established in the Regula-
tions of the Council for the Approximation of Legislation for the member countries regard-
ing electromagnetic compatibility (2004/108/EG). Additional fault: <1% 