5 Start Operation
Connect sensor, turn on device via key.
After segment test the device displays some configuration:
• If the logger function is not off the time of the integrated clock will shortly be displayed.
If a zero point adjustment was carried out the display shows shortly „nuLL Corr“.
After changing the battery the clock-setting menu is activated automatically (‚CLOC‘). Check the clock
and adjust, if necessary (p.r.t. chapter 6).
After that the device is ready for measuring.
6 Configuration
To change device settings, press Menu (key 4) for 2 seconds. This will call the configuration menu (main
display: „SEt“).
Pressing key Menu changes between the menus, pressing (key 3) jumps to the referring parameters,
which can be selected with key (key 3).
The parameters can be changed with (key 2) or (key 5).
Pressing Menu again jumps back to the main configuration menu and saves the settings.
Quit (key 6) finishes the configuration and returns to standard measuring operation.
Please note: All here specified values are applying for all devices if there is no explicit mark.
The values for the different types of device can be checked up in chart (page 8)
Menu
Param.
Values
Description
C_log
C_dat
Stor
‚Menu‘
or
SEt
ConF
Set Configuration: Generic Configurations
Unit
mbar,bar..
Unit: Unit of display
*
*
SL
oFF/on
Sea level correction: on or off (only available at GMH 3181-12)
*
*
Alti
-2000..9999
Altitude: Input of altitude above sea level [m] (only if SL=on)
*
*
rAtE
Rate: Measuring rate (p.r.t. chapter 7.1)
*
*
Slo
Slow measuring rate (4Hz filtered, low power consumption)
*
*
FASt
Fast measuring rate, filtered (>1000Hz)
*
*
P.dEt
Peak detection: fast measuring rate, unfiltered (>1000Hz)
*
*
t.AVG
1-120
Averaging period in seconds, used by the averaging function
*
*
oFF
Averaging function deactivated
*
*
P.oFF
1-120
Auto Power Off time in minutes
oFF
Auto Power Off deactivated
Out
oFF
Function of the output: No output function, lowest power consumption
SEr
Output is serial interface
dAC
Output is analogue output 0...1V
Adr.
01,11..91
Base address of interface
dAC.0
p.r.t. chart
Enter desired value at which the analogue output potential should be 0V
(if Out = dAC)
dAC.1
p.r.t. chart
Enter desired value at which the analogue output potential should be 1V
(if Out = dAC)
SEt
CAL
Set Calibration: Adjustment of Sensor
*
OFFS
p.r.t. chart
The offset of sensor will be displaced by this value to compensate for
deviations in the probe or in the measuring device.
*
oFF
Zero displacement inactive (=0.00)
*
SCAL
-2.000 ...
2.000
The measuring scale of sensor will be changed by this factor [%] to
compensate deviations of temperature probe or measuring device
*
oFF
Scale correction factor inactive (=0.000)
*
1 2 3
4 5 6
ON
OFF
max
Store
CAL
min
Set
Menu Quit